Dual-Scale Display Feature
The ProtEX-MAX PD8-6000 has a rather unique, and very flexible dual-scale capability. This is of particular value in
level applications where a second scaled display can represent the measured input in a different form (i.e. gallons &
height). Both displays are independently scaled and are based on the 4-20 mA input signal. Beyond level, this
function has been used for pressure & force, current & power, feet & meters, GPM & CFM, and more.
Gallons & mA
Gallons & Height
Gallons & Percent
Gallons & Head PSI
Rounding for Even Steadier Display
The rounding feature is used to give the user a
steadier display with fluctuating signals. It causes the
display to round to the nearest value according to the
rounding value selected (1, 2, 5, 10, 20, 50, or 100).
For example, with a rounding value of 10, and an
input of 12346, the display would indicate 12350.
Other Uses for Second Line
The second line can also be used indicate units, a
tag, or even a setpoint as the following pictures
illustrate:
Volume in Tank
Gallons & Setpoint
Level in Feet
Pressure Indication