PD663 Explosion-Proof Loop-Powered Meter
Instruction Manual
10
Calibrating the Meter (
Cal
)
To scale the meter without a signal source refer to
The meter can be calibrated to display the process in
engineering units by applying the appropriate input
signal and following the calibration procedure.
The use of a calibrated signal source is strongly
recommended.
Press the
Up
arrow button to scroll to the
Calibration
menu (
CAL
) and press
Enter
.
If using linear signal input conditioning, enter the
number of calibration points (2-32).
The meter displays
in1
. Apply a known signal and
press
Enter
. The display will flash while accepting the
signal.
When the meter displays
d1
, press
Enter
. Enter a
corresponding display value for the signal input, and
press
Enter
to accept.
The meter displays
in2
. Apply a known signal and
press
Enter
. The display will flash while accepting the
signal.
When the meter displays
d2
, press
Enter
. Enter a
corresponding display value for the signal input, and
press
Enter
to accept.
Re-calibrating the Internal Calibration
Reference (
ICal
)
The
Internal Calibration
(
ICAL
) menu, located in the
Advanced
features menu, is used to
recalibrate the internal calibration reference.
Recalibration is recommended at least every twelve
months. Refer to
for instructions.
Advanced Features Menu
To simplify the setup process, functions not needed
for most applications are located in the
Advanced
features menu. Press and hold the
Menu
button for
five seconds to access the
Advanced
features menu.
Advanced Features Menu & Display
Messages
The following table shows the
Advanced
features
menu functions and messages in the order they
appear in the menu.
Display
Parameter
Action/Setting
Fnc
Input Function
Set linear or square
root input conditioning
function
Lnr
Linear
Set linear scaling
Sqr
Square Root
Set square root input
conditioning function
info
Information
Enter the
Information
menu
Sft
Software
Information
Software release
number
ver
Version
Meter firmware version
C
Calibration
Temp (
°
C)
Temperature at time of
I-calibration (
°
C)
F
Calibration
Temp (
°
F)
Temperature at time of
I-calibration (
°
F)
FIL
Filter
Set filter function level
ICAL
I-Calibration
Internal master factory
calibration
rst
Reset Defaults
Restore factory default
parameter settings