6
5335V116-UK
Electrical specifications, input:
Max. offset ................................................................. 50% of selec. numerical max. value
RTD and linear resistance input:
Cable resistance per wire (max.) ....................... 5
Ω
(up to 50
Ω
per wire is possible with reduced measurement accuracy)
Sensor current ........................................................... Nom. 0.2 mA
Effect of sensor cable resistance (3- / 4-wire) .... < 0.002
Ω
/
Ω
Sensor error detection ........................................... Yes
Short circuit detection ........................................... If 0% > 30
Ω
TC input:
Cold junction compensation ................................ < ±1.0°C
External CJC with Ni100 or Pt100 ................... -40
≤
Tamb.
≤
135°C
Sensor error detection ........................................... Yes
Sensor error current:
When detecting
............................................... Nom. 33
μ
A
Else ....................................................................... 0
μ
A
Short circuit detection ........................................... No
Type
Min.
temperature
Max.
temperature
Min.
span
Standard
B
E
J
K
L
N
R
S
T
U
W3
W5
+400°C
-100°C
-100°C
-180°C
-100°C
-180°C
-50°C
-50°C
-200°C
-200°C
0°C
0°C
+1820°C
+1000°C
+1200°C
+1372°C
+900°C
+1300°C
+1760°C
+1760°C
+400°C
+600°C
+2300°C
+2300°C
100°C
50°C
50°C
50°C
50°C
50°C
100°C
100°C
50°C
50°C
100°C
100°C
IEC584
IEC584
IEC584
IEC584
DIN 43710
IEC584
IEC584
IEC584
IEC584
DIN 43710
ASTM E988-90
ASTM E988-90
RTD
type
Min.
value
Max.
value
Min.
span
Standard
Pt100
Ni100
Lin. R
-200°C
-60°C
0
Ω
+850°C
+250°C
7000
Ω
10°C
10°C
25
Ω
IEC 60751
DIN 43760
-----