Troubleshooting
If the Dial Flowmeter fails to function, consult the Trouble Shooting
Guide below. If the problem cannot be solved, consult your provider.
Problem
Probable Cause
Remedy
No Flow
•
Flowmeter set in OFF
position
•
Dial set between settings
• Internal failure
• Hoses are not properly
connected between the
DISS inlet and Flowmeter
•
Turn to desired setting
•
Move Dial to flow
setting
•
Return to Powerex
•
Remove Front Panels
and check connections
Dial will not turn /
Hard to turn
•
Dial seized
•
Return to Powerex
Will not shut off
•
Flowmeter not in OFF
position
•
Turn Dial to the OFF
position
To prevent risk of burns, fire, or injury to person(s):
•
ALWAYS confirm prescribed flow before administering to patient and
monitor flow on a frequent basis.
•
Always follow ANSI and CGA standards for Medical Gas Products and
Flowmeters and Oxygen Handling.
•
DO NOT use oils, greases, organic lubricants or any combustible
materials on or near this Dial Flowmeter.
•
DO NOT use near any type of flame or flammable/explosive substances,
vapors or atmosphere.
•
DO NOT smoke in an area where oxygen is being administered.
•
This Dial Flowmeter contains magnetic, ferrous material that may affect
the results of an MRI.
•
NO FLOW is delivered when the Indicating Pointer
▼
is pointing to OFF.
•
Flow is available only at the listed increments. There is NO FLOW
between increments.
•
Only personnel instructed and trained in its use
should operate this Dial Flowmeter.
•
Be sure all connections are tight and leak free.
•
Only use oxygen-safe leak detector to test for leaks.
•
DO NOT autoclave.
•
DO NOT gas sterilize with (EtO) Ethylene Oxide.
•
DO NOT clean with aromatic hydrocarbons.
•
DO NOT immerse Dial Flowmeter in any kind of liquid. This will void the
warranty.
•
DO NOT attempt to repair the Dial Flowmeter.
•
Store the Dial Flowmeter in a clean area when not in use.
NOTE:
The medical oxygen outlet you are connecting to in the
patient room must be labeled “Not For Use With A Flowmeter”.
Flow regulation is controlled by the Dial Flowmeter that is on the
Sleep Lab Panel.