Sleep Lab Panel

Powerex  

  150 Production Drive  

  Harrison, OH  45030  

  USA

P   1.888.769.7979  

  F   513.367.3125  

  www.powerexinc.com

PX-99-0534  

  03/2017

Page 4 of 8

Troubleshooting

If the Dial Flowmeter fails to function, consult the Trouble Shooting 

Guide below. If the problem cannot be solved, consult your provider.

Problem

Probable Cause

Remedy

No Flow

• 

Flowmeter set in OFF 

position

• 

Dial set between settings

• Internal failure

• Hoses are not properly 

connected between the 

DISS inlet and Flowmeter

• 

Turn to desired setting

• 

Move Dial to flow 

setting

• 

Return to Powerex

• 

Remove Front Panels 

and check connections

Dial will not turn / 

Hard to turn

• 

Dial seized

• 

Return to Powerex

Will not shut off

• 

Flowmeter not in OFF 

position

• 

Turn Dial to the OFF 

position

To prevent risk of burns, fire, or injury to person(s):

• 

ALWAYS  confirm  prescribed  flow  before  administering  to  patient  and 

monitor flow on a frequent basis.

• 

Always follow ANSI and CGA standards for Medical Gas Products and 

Flowmeters and Oxygen Handling.

• 

DO  NOT  use  oils,  greases,  organic  lubricants  or  any  combustible 

materials on or near this Dial Flowmeter.

• 

DO NOT use near any type of flame or flammable/explosive substances, 

vapors or atmosphere.

• 

DO NOT smoke in an area where oxygen is being administered.

• 

This Dial Flowmeter contains magnetic, ferrous material that may affect 

the results of an MRI.

• 

NO FLOW is delivered when the Indicating Pointer 

▼ 

is pointing to OFF.

• 

Flow  is  available  only  at  the  listed  increments.  There  is  NO  FLOW 

between increments.

• 

Only personnel instructed and trained in its use 

should operate this Dial Flowmeter.

• 

Be sure all connections are tight and leak free.

• 

Only use oxygen-safe leak detector to test for leaks.

• 

DO NOT autoclave.

• 

DO NOT gas sterilize with (EtO) Ethylene Oxide.

• 

DO NOT clean with aromatic hydrocarbons.

• 

DO NOT immerse Dial Flowmeter in any kind of liquid. This will void the 

warranty.

• 

DO NOT attempt to repair the Dial Flowmeter.

• 

Store the Dial Flowmeter in a clean area when not in use.

NOTE: 

The  medical  oxygen  outlet  you  are  connecting  to  in  the 

patient room must be labeled “Not For Use With A Flowmeter”. 

Flow regulation is controlled by the Dial Flowmeter that is on the 

Sleep Lab Panel.

Summary of Contents for PX-SLP-1

Page 1: ...y to service layout design Flowmeter control panel incorporates dial adjustment Ease of maintenance DISS gas specific connections Single flowmeter panel Dual flowmeter panels PX SLP 1 Sleep Lab Panel PX SLP 2 Sleep Lab Panel Contents Features Benefits 1 Introduction 2 Major components 3 Installation 5 Installation of Inlet Outlet DISS 5 Sleep Lab Panel Installation 5 Maintenance 6 Flowmeters 7 Ope...

Page 2: ...DISS assemblies must be installed in accordance with Installation of Inlet and Outlet DISS Assemblies above Sleep Lab Panel Installation 1 Powerex Sleep Lab Panels may be plumbed in either direction with the patient use side on the left or the right This saves the plumber time and money 2 The rough in back box is shipped with a cardboard dust cover installed You will need to remove the dust cover ...

Page 3: ...Operating Temperature 69 F 20 6 C to 71F 21 6 C Storage Temperature 40 F 40 C to 140 F 60 C Max Humidity 95 Note Storage Transport outside the specified range may cause damage to the flow meter Specifications are subject to change without notice Operating Instructions Note Tri Tech Medical Inc recommends the use of kink proof patient tubing 1 Be sure Dial Flowmeter is in the OFF position 2 Connect...

Page 4: ...Handling DO NOT use oils greases organic lubricants or any combustible materials on or near this Dial Flowmeter DO NOT use near any type of flame or flammable explosive substances vapors or atmosphere DO NOT smoke in an area where oxygen is being administered This Dial Flowmeter contains magnetic ferrous material that may affect the results of an MRI NO FLOW is delivered when the Indicating Pointe...

Page 5: ...e Harrison OH 45030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com PX 99 0534 03 2017 Page 5 of 8 Appendix A Technical Specifications NOTE Sleep Lab Panels me be plumbed in either direction with the patient use side on left or the right ...

Page 6: ...Sleep Lab Panel Powerex 150 Production Drive Harrison OH 45030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com PX 99 0534 03 2017 Page 6 of 8 Notes ...

Page 7: ...es licenses or other fees imposed by any governmental authority upon the production sale shipment and or use of Products covered hereunder The warranties and remedies herein are the sole and exclusive remedy for any breach of warranty or for any other claim based on any Defect or non performance of the Products whether based upon contract warranty or negligence d Coverage The warranty provided her...

Page 8: ...of Products for certain purposes which may vary from those in neighboring areas While Powerex attempts to assure that its Products comply with such codes it cannot guarantee compliance and cannot be responsible for how the product is installed or used Before purchase and use of a Product please review the Product applications and national and local codes and regulations and be sure that the Produc...

Reviews: