background image

Sleep Lab Panel

Powerex  

  150 Production Drive  

  Harrison, OH  45030  

  USA

P   1.888.769.7979  

  F   513.367.3125  

  www.powerexinc.com

PX-99-0534  

  03/2017

Page 3 of 8

11. 

When  all  connections  are  complete  and  leak  tested  with  an 

oxygen compatible leak test solution, fasten the sleep lab front 

panel to the back box with the four #8-32 x ¾” screws provided.

12. 

On the front panel, label each flowmeter in the space provided 

with  the  sleep  room  it  controls  (label  not  provided).  See 

Appendix A.

Mounting Screws

Miss-connection of the gases could lead to serious 

or  fatal  injury  to  patients.  Following  installation, 

DISS  connections  must  be  tested  for  cross-connection  (per  appropriate 

standards) to ensure that the intended services are correctly connected to 

the appropriate service lines.

Make  certain  the  labeling  coincides  with  the  gas 

service,  and  areas  controlled  by  the  Sleep  Lab 

Panel and that it is easily read.

Maintenance

1. 

Clean the exterior of the stainless steel Sleep Lab Panel with a 

glass cleaner. Strong solvents will damage the etched screened 

printing on the sleep lab panel.

2. 

Never use glass cleaner on the flowmeters.

3. 

If it is necessary to clean the flowmeters wipe them clean with 

a cloth dampened with water

4.  Wipe dry with a clean dry cloth.

Flowmeters

Accuracy: ± 10% of the indicated flow.
Operating Temperature: 69°F (20.6°C) to 71F° (21.6° C)
Storage Temperature: -40°F (40°C) to 140°F (60° C)
Max Humidity: 95 %

Note: 

Storage / Transport outside the specified range may cause 

damage to the flow meter.

Specifications are subject to change without notice.

Operating Instructions

Note: 

Tri-Tech  Medical,  Inc.  recommends  the  use  of  kink  proof 

patient tubing.

1. 

Be sure Dial Flowmeter is in the “OFF” position.

2. 

Connect the Dial Flowmeter to a 50 psi (3.4 bar) Oxygen Source.

3. 

Set the flow by aligning the Indicating Pointer 

 to the desired 

flow.

Note: 

The accuracy of flow will be adversely affected by any variation 

of inlet pressure, ambient temperature or outlet resistance from 

rated values.

Read  this  user  manual  before  installing  or 

operating the Dial Flowmeter.

There is NO FLOW between settings. The Indicating Pointer 

▼ 

must point 

to a specific number on the dial to obtain flow.

ONLY  use  appropriate  indexed  fittings  to  connect 

the  Dial  Flowmeter  to  appropriate  source.  Use 

Oxygen connections for oxygen flowmeters and use Air connections for Air 

flowmeters.

Cleaning/Maintenance

1. 

Disconnect all connections before cleaning.

2. 

Clean  exterior  surfaces  of  the  Dial  Flowmeter  with  a  cloth 

dampened with water.

3.  Wipe dry with a clean cloth.
4. 

Store the Dial Flowmeter in a clean area free from grease, oil 

and other sources of contamination.

DO NOT immerse the Dial Flowmeter in any kind 

of  liquid.  This  will  cause  damage  to  the  Dial 

Flowmeter and will void the warranty.
This Dial Flowmeter is not field serviceable.

Summary of Contents for PX-SLP-1

Page 1: ...y to service layout design Flowmeter control panel incorporates dial adjustment Ease of maintenance DISS gas specific connections Single flowmeter panel Dual flowmeter panels PX SLP 1 Sleep Lab Panel PX SLP 2 Sleep Lab Panel Contents Features Benefits 1 Introduction 2 Major components 3 Installation 5 Installation of Inlet Outlet DISS 5 Sleep Lab Panel Installation 5 Maintenance 6 Flowmeters 7 Ope...

Page 2: ...DISS assemblies must be installed in accordance with Installation of Inlet and Outlet DISS Assemblies above Sleep Lab Panel Installation 1 Powerex Sleep Lab Panels may be plumbed in either direction with the patient use side on the left or the right This saves the plumber time and money 2 The rough in back box is shipped with a cardboard dust cover installed You will need to remove the dust cover ...

Page 3: ...Operating Temperature 69 F 20 6 C to 71F 21 6 C Storage Temperature 40 F 40 C to 140 F 60 C Max Humidity 95 Note Storage Transport outside the specified range may cause damage to the flow meter Specifications are subject to change without notice Operating Instructions Note Tri Tech Medical Inc recommends the use of kink proof patient tubing 1 Be sure Dial Flowmeter is in the OFF position 2 Connect...

Page 4: ...Handling DO NOT use oils greases organic lubricants or any combustible materials on or near this Dial Flowmeter DO NOT use near any type of flame or flammable explosive substances vapors or atmosphere DO NOT smoke in an area where oxygen is being administered This Dial Flowmeter contains magnetic ferrous material that may affect the results of an MRI NO FLOW is delivered when the Indicating Pointe...

Page 5: ...e Harrison OH 45030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com PX 99 0534 03 2017 Page 5 of 8 Appendix A Technical Specifications NOTE Sleep Lab Panels me be plumbed in either direction with the patient use side on left or the right ...

Page 6: ...Sleep Lab Panel Powerex 150 Production Drive Harrison OH 45030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com PX 99 0534 03 2017 Page 6 of 8 Notes ...

Page 7: ...es licenses or other fees imposed by any governmental authority upon the production sale shipment and or use of Products covered hereunder The warranties and remedies herein are the sole and exclusive remedy for any breach of warranty or for any other claim based on any Defect or non performance of the Products whether based upon contract warranty or negligence d Coverage The warranty provided her...

Page 8: ...of Products for certain purposes which may vary from those in neighboring areas While Powerex attempts to assure that its Products comply with such codes it cannot guarantee compliance and cannot be responsible for how the product is installed or used Before purchase and use of a Product please review the Product applications and national and local codes and regulations and be sure that the Produc...

Reviews: