background image

 

SECTION 1 

INTRODUCTION AND DESCRIPTION 

 

1.1

 

 The ST-375 (P/N 37512) High Rate Discharge Unit is For testing SLI (Starting, Lighting, Ignition) engine start    

       batteries. It is used to accept or reject reject 12-Volt engine start batteries which have been fully charged prior to   
       testing.  The ST-375 is designed to work on batteries that have  CCA (Cold Cranking Amps) ratings in the  
       approximate range of 700 CCA to 1,000 CCA.  However it will also accurately test batteries with a higher CCA  
       rating such as those used in large commercial trucks and military tactical vehicles.  The unit is essentially an  
       SAE (Society of Automotive Engineers) tester which places a 375 Ampere load on the battery for a period of 15  
       seconds. A digital voltmeter on the unit reads and holds the battery voltage at the end of 15 seconds. If the  
       voltage is at or above 10.0 volts, the battery is ready for issue. If the voltage is below 10.0 volts the battery is  
       rejected. The battery may be recharged for another test depending on circumstances such as if this were a first  
       attempt to condition the rejected battery.  The ST-375 has a high wattage resistive load bank. The resistors do  
       not glow during the 15 second test. The control switch has an automatic position which is pressed, released, and  
       tests for 15 seconds. A manual position can be pressed for less then 15 seconds. An internal 500 Ampere Shunt  
       has its test terminals (Kelvin Terminals) connected to a pair of test points on the front panel. The voltmeter also  
       has test points. 
 
1.2   This procedure describes the verification of accuracy of Power Products High rate Discharger Tester P/N 

37512. The 37512 being verified is referred to herein as the TI (Test Instrument). The TI operates from the 
battery being tested and does not require any other power. 

 
1.3

 

This procedure includes tests of essential performance parameters only. Any malfunction or out of limit 

readings noticed during these procedures, whether specifically tested for or not, should be referred to power 
products. 

 

Table 1. Verification Description 

 

TI Characteristics 

Performance Specification 

Test Method 

Voltmeter 
 
 
 
Internal shunt and load current 
(NOTE: Shunt is 500Amp. 
50 Millivolt)  
 
  
 
 
 
 
2 internal timers. Timer 1 
holds the voltmeter reading at 
the end of 15 second test.  
 

 

Timer 2  disconnects load 3 
seconds after voltage hold (18 
seconds total) 

Range: 0 to 13 V (Meter range 99.9 V) 
Tolerance: 0.2 V 
 
 
Range: 0 to 38MV/380Amp 
Tolerance:  1.0 Millivolts/ 10 Amps 
Actual load tolerance is 375 Amps ± 10 
Amps at 10.5 Volts 10 Seconds of load 
(.028 Ohms at 10 Seconds) 
 
 
 
 
TIME: tolerance 2 Seconds 
 
 
 
 
TIME: tolerance 2 Seconds 

Connect to battery or power supply, keep 
TI power switch OFF. Compare TI 
voltmeter reading with multi-meter. 
 
 Load the TI with a battery or high current 
12 volt power supply.  Use external shunt 
and 2 multi-meters.  Compare TI reading 
with one multi-meter and with external 
shunt and second multi-meter. 
NOTE: Amp reading is 10X millivolt 
reading 
 
 
Check with Quartz stop-watch or timer. 
 
 
 
 
Check with Quartz stop-watch or timer. 

 
 
 
 
 

Summary of Contents for ST-375

Page 1: ... 375 905 2 TECHNICAL MANUAL INSTRUMENT VERIFICATION PROCEDURE HIGH RATE DISCHARGE TESTER POWER PRODUCTS P N 37512 MODEL ST 375 MFD BY POWER PRODUCTS INC 27 PAMARON WAY STE E NOVATO CA 94949 CAGE CODE 3C8X4 www power products com POC PAUL CHAIT ...

Page 2: ...ions 2 4 Verification Process 2 4 ILLUSTRATIONS Figure Page 1 Voltmeter Test Setup 3 2 Ammeter Test Setup 3 3 Timing Setup Test 5 Data Sheet 375 9 05 TABLES Table Page 1 Verification Description 1 2 Equipment Requirements 2 3 Tolerance Limits 3 4 Timing Requirements 4 5 Summary Table of Tolerance Requirements 5 ...

Page 3: ... Terminals connected to a pair of test points on the front panel The voltmeter also has test points 1 2 This procedure describes the verification of accuracy of Power Products High rate Discharger Tester P N 37512 The 37512 being verified is referred to herein as the TI Test Instrument The TI operates from the battery being tested and does not require any other power 1 3 This procedure includes te...

Page 4: ...re Table 2 Equipment Requirements for TI Item Minimum Use Specifications Calibration Equipment SECTION 3 PRELIMINARY OPERATIONS 3 1 Ensure that power switch is set to OFF to avoid arcing or damage to the battery or power supply terminals or the TI battery clamps 3 2 Connect the Battery or power supply to the TI Observe polarity SECTION 4 VERIFICATION PROCESS NOTE Unless otherwise specified verify ...

Page 5: ... TI DC BATTERY CABLE TO FROM THE BATTERY 4 1 6 LOAD CURRENT TESTS Connect the equipment as shown in Figure 2 4 1 7 Ensure that the AUTO OFF MANUAL switch is set to OFF 4 1 8 Connect shunt as described in table 2 item 2 2 and power supply or optima battery as described in table 2 item 2 3 4 1 9 Set switch to Manual to obtain Readings Load will not remain on over 15 seconds Refer to table 5 par 4 1 ...

Page 6: ...to AUTO The green led illuminates 4 1 14 At the end of 15 seconds the voltmeter holds freezes its voltage Approximately 3 seconds later a contactor click out is heard and the green led turns off Table 4 Timing Requirements TI Time Tolerance Limit AUTO test led ON Load ON START Voltmeter Holds Battery End Voltage 15 Sec Within 2 Sec Retains the voltage reading on hold Load contactor drops out Green...

Page 7: ...5V 75 375 Amp XAMPLE Actual voltage reading 10 9 difference is 0 4 Amps ptable Test reading is Amperes 4 If actual volts are 10 1 differences is 0 4 le ps 10 5V S 1 2 Multiply Difference by 35 3 If higher than 10 5 add to 3 4 If lower than 10 5 subtract from 5 Acceptable Result must be the actual reading 10 Amperes E STEPS 1 2 0 4 X 35 Amps 14 Amps 3 14 Amps 375 Amps 389 4 5 Acce 389 10 Amps 379 t...

Page 8: ...sitive Black to Negative Meter reads Battery Open Circuit voltage Toggle switch to AUTOMATIC this is a 15 sec load test At end of 15 seconds meter locks in voltage reading Relay click heard after lock in If voltage is at or above 10 0 volts Battery is acceptable Below 10 0 volts is unacceptable For Manual test press and hold switch down Cuts Off at 15 sec Set switch to center OFF before removing b...

Reviews: