Electromagnetic Compatibility (EMC)
Precautions
The Pro7M model is suitable for use in
commercial medical establishments,
including those directly connected to
the public low-voltage power supply
network that supplies buildings used for
domestic purposes. Both models have
been tested against, and have passed
the applicable requirements of relevant
electromedical standards, including EN
60601-1-2:2007.
The Pro7M model emits electromagnetic
energy to perform its intended function.
Nearby electronic equipment may be
affected by this emission. Similarly
the Pro7M model may be affected by
electromagnetic emissions from other
equipment in the vicinity.
The Pro7M model is intended for use
in the electromagnetic environment
specified below. The customer or user of
the Pro7M model should ensure that it is
used in such an environment:
– Floors should be wood, concrete or
ceramic tile. If the floor is covered with
synthetic material, the relative humidity in
the room should be at least 30%.
– Power quality should be that of a typical
domestic, commercial or hospital
environment as appropriate.
– The Pro7M model should not be used
during power interruptions.
– Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
The Pro7M model is intended for use in an
electronic environment in which radiated
radio-frequency (RF) disturbances
are controlled.
The customer or user of the Pro7M model
can help prevent electromagnetic
interference by maintaining a minimum
distance between portable and
mobile RF communications equipment
(transmitters) and the Pro7M model
as recommended in this user manual,
according to the maximum output power
of the communications equipment:
The adequate separation distance
to provide base immunity to RF
disturbances is 1.0m.
•
USB Port - USB port supplies 5Volt DC
current. This port is only used for soft-
ware updates and uploading personal
pictures to user profiles. The operator
must take care to not touch the USB port
and a patient simultaneously.
•
LAN port - the Pro7M has included a
LAN port on the lower back side of
the machine. This LAN port is used
to connect a LAN cable to access the
internet.The operator must take care to
not touch the LAN port and a patient
simultaneously.
•
Phone jack adapter- On the Control
Panel of the Pro7M, there is a phone
jack adapter for sound playback. The
operator must take care to not touch
the Phone jack adapter and a patient
simultaneously.
Retain this user manual for your future
reference.