![Posey Posey SITTER II Instruction Manual Download Page 20](http://html1.mh-extra.com/html/posey/posey-sitter-ii/posey-sitter-ii_instruction-manual_1582161020.webp)
20
Sitter II
®
Monitoring with a Sensor
The following instructions will help you set up and safely use the Sitter II.
TIPS TO PROTECT SENSORS FROM DaMagE
To avoid inconvenience to staff and patients, and to protect sensors from damage, you
should follow these steps:
• only use Posey sensors with the Posey alarm.
• When routing sensor cord to alarm, check that there is no stress on cord. Cord
must be clear of all moving parts of bed or chair to prevent sensor failure.
•
NEVER
jerk or pull on the cord to remove RJ11 plug. Doing so will
damage cord wires or plug.
•
alWayS
use the plastic tab to release plug (fig. 19).
FaIlSaFE FEaTuRE
The Posey Sitter II contains a “failsafe” feature that activates the Posey alarm if the sensor
is removed from the alarm when the power is on.
FOR SaFE uSE IN all SENSOR MODES
To reduce the risk of serious injury or death, alWayS follow these steps after putting the
sensor in place and before leaving patient unattended (see instructions below). DO NOT use
any alarm or sensor that does not alarm each time it is tested.
1. Make sure alarm is ON and in monitoring mode (lED light is flashing green above POWER
button).
2. Check that the RJ11 plug on the sensor cable is not damaged (plug broken, or wires
disconnected) and is securely connected to the alarm.
3. Disconnecting the sensor from the alarm when the power is on will cause the alarm to
activate. This is called a “failsafe” mode. Disconnect the sensor to make sure the failsafe
mode works. DO NOT use the alarm if the alarm does not sound when the sensor is
disconnected.
4. If in use, check that the nurse call cable is securely connected to the alarm and the nurse
call panel. alWayS test alarm and nurse call function if ‘nurse call’ cable is plugged into
the alarm and wall jack. activate the alarm (remove pressure from sensor, unfasten chair
belt sensor, or activate PIR sensor) and make sure the nurse call light for the proper bed and
room activate in the appropriate nurse’s station location. Remove the cable from the wall
jack and make sure the visual or audible alert at the nurse’s station immediately activates.
5. Inspect sensor cord and nurse call cable (if in use) to ensure they are out of the footpath
and DO NOT pose a tripping hazard.
Fig. 19
Plastic Tab
Summary of Contents for Posey SITTER II
Page 35: ...Notes ...