9.4.1 Configuring the relays in mode 1
In the example below, the relays are configured thus:
Use the
and
keys to
increase and decrease the values.
Use the
and
keys to move
from selection to selection.
Once the two relays are correctly configured, press the
ENTER key.
CONFIG
MODE 1: ADJUSTEMENT OF THE
THRESHOLDS AND HYSTERESIS
Rel 1 connected to pH measurement:
Alarm threshold:12.01 pH Hyst : 0.10 pH
Above the threshold, R1 is on
Rel 2 connected to Temperature:
Alarm threshold:20.01 °C Hyst : 0.06 °C
Above the threshold, R2 is on
CONFIG
WHAT DO YOU WANT TO DO?
Graphic screen
4-20mA output adjustment
Relays outputs adjustment
Measure unit
Language
Factory default value reminder
Esc: Cancel the procedure
Enter: Validate the choice
12.01
11.91
R1
20.01
20.94
R2
pH
Temperature