Once the 4mA outputs are correctly adjusted, press the
ENTER key.
Use the
and
keys to increase
and decrease the 20mA value.
Use the
and
keys to select
either the 4-20mA output for the pH
or the 4-20mA output for the
temperature.
CONFIG
4-20mA OUTPUTS
GENERATION DU 20mA
Enter the value
read on the
output 4-20mA pH
20.00
mA
Enter the value read
on the output 4-20mA
temperature
20.00
mA
Once the 20mA outputs are correctly adjusted, press
the ENTER key.
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Graphic screen
4-20mA output adjustment
Relays outputs adjustment
Measure unit
Language
Factory default value reminder