User manual
PONSEL
Reference: NOTICE_ACTEON_2010_v001-UK
V: 001
23/56
6.1.2 Adjusting the slope of the pH sensor
If neede, you can adjust the slope of your sensor by implementing the following procedure:
CALIBRATION
SENSORS CALIBRATION
SENSOR pH
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to
select the type of calibration
and sensor to be calibrated.
Select the
Gain adjust
menu then press the
ENTER key.
Use the
and
keys to adjust
the value of the buffer (from
4.00 to 10.98).
To validate the gain adjustment, press the ENTER
key.
CALIBRATION
OF THE SENSOR pH
Immerse sensor in
a pH 4.01 buffer
Standard :
Measure
:
pH
pH
ESC: Leave the current gain
ENTER: Validate the new gain
4.01
4.20
WAIT, Nonstable measurement
MEASUREMENT STABILISATION.
Immerse the sensor in the pH 4.01 or 10.01 buffer solution.