Guidance and manufacturer’s declaration-electromagnetic immunity
The MBPN002, MB0500300 is intended for use in the electromagnetic environment specified below.
The customer or the user of the MBPN002, MB0500300 should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply lines
+ 1kV for input/output
lines
+ 2kV for power
supply lines
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
+ 1kV line(s) to line(s)
+ 2kV line(s) to earth
+ 1kV differential
mode
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 sec.
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 sec.
Mains power quality should be
that of a typical commercial
or hospital environment. If
the user of the MBPN002,
MB0500300
requires continued
operation during power
mains interruptions, it is
recommended that the
MBPN002, MB0500300 be
powered from an uninterruptible
power supply or a battery.
Power frequency(50,
60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
The MBPN002, MB0500300
power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.