displayed bath temperature is 20.0°C, an
offset of 0.2 displays a value of 20.2°C. An
offset of -0.2°C will display 19.8°C. Range is
-5°C .. +5°C. Press page key to confirm
setting when altered.
note:
When tune is set to on, PID values can not
be set. Change tune to off en alter PID
values.
Press page key to confirm setting when
altered.
7.6 Quick start
To quickly start operating the bath:
•
Fill the bath with sufficient fluid to submerse the
cooling coil. However do not overfill the bath fully as
the fluid will exit through the overflow outlet,
•
Place the power plug,
•
Switch the bath on using the mains switch,
•
Choose a working temperature (set point):
Press once. Display will indicate "SP.1". Alter set point
temperature by using up and down key to select the
desired bath temperature.
Press to display bath temperature again (or wait 10 seconds).