Page 63/68
Emissions Test
Compliance
Electromagnetic environment - Guidance
RF Emission CISPR 11
Group 1
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF Emission CISPR 11
Class B
The device is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions IEC
61000-3-2
Class B
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Class B
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment -
Guidance
ESD IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material the relative
humidity should be at
least 30 %.
Electric fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 2 kV for power supply
lines
Mains power supply
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power supply
should be that of a
typical commercial or
hospital environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-
11
0% 0.5 periods 0°
40% 5 periods 0°
70% 25 periods 0°
0% 250 periods 0°
0% 0.5 periods 0°
40% 5 periods 0°
70% 25 periods 0°
0% 250 periods 0°
Mains power supply
should be that of a
typical commercial or
hospital environment.
Power Frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.