Page 50/68
4.2 Trouble-shooting guide
Vast majority of service queries relate to malfunctions when switching device on, malfunctions
when using touch screen or error messages when performing a measurement. It is only in rare
cases that these malfunctions are caused by broken hardware. The following chapters provide
a step-by-step guide for trouble-shooting.
4.2.1 Trouble-shooting when switching device on
a) The device cannot be switched on
If the device cannot be switched on, connect the power adapter to a plug and check
whether the green LED lights up on the upper side of the power adapter. Now connect the
12V charging cable to the device. If the device now switches on, check whether
rechargeable batteries are inserted (see chapter 2.3
”).
b) The device switches off again immediately
If the device switches off again immediately, the rechargeable batteries are likely to be
discharged. Connect the 12V charging cable to charge the batteries. Check whether the
power supply unit is receiving current (green LED on top of the power adapter lights up).
You can use the device while it is being charged. To do so, simply switch the device on
while the 12V charging cable is connected.
4.2.2 Malfunctions when using touch screen
a) Screen switches off suddenly
To save power, the device automatically switches the screen off after some time. Touch the
screen to activate it again. If the screen is not activated by touching it, the device has
switched off automatically. Simply switch it back on again by pressing the On/Off button.
Define in Basic settings after how many seconds the device should automatically
switch off the screen and after how many minutes it should automatically switch off
itself (see chapter 4.1.2