90
Meter detects blood glucose
level is lower than 20 mg/dL
(1.1 mmol/L).
Indicates low blood glucose
level. Repeat test. If result
still occurs, consult your
physician immediately.
Temperature is high during
the test procedure.
Result may be larger
variation. Move to a cooler
environment (10 - 40˚C or 50
- 104˚F) and wait 15 minutes
before re-testing.
Temperature is low during
the test procedure.
Result may be larger
variation. Move to a warmer
environment (10 - 40˚C or 50
- 104˚F) and wait 15 minutes
before re-testing.
ACTION
DISPLAY
WHAT IT MEANS
Summary of Contents for pM-100 fasTTest