Comparing Control Solution Results
The plusmed fasTTest Blood Glucose Monitoring System is functioning properly if the
results fall inside the specified range printed on the Test Strip vial. If the test results fall
outside the specified range, repeat the test.
Results falling outside the specified range may be caused by
• Error in performing the test,
• Control solution temperature is lower than 15°C (59°F)
or higher than 35°C (95°F),
• Expired or contaminated Control Solution,
• Expired or contaminated Test Strips,
• Meter malfunction.
Note:
The result will not be incorporated into the average in the Control Solution test
“ ctl “ mode.
DO NOT
use the system if the problem persists.
Please contact our central technical service.