Planmeca ProSensor 5
PLANMECA PROSENSOR CONTROLBOX INDICATOR LIGHT
User’s Manual
6
PLANMECA PROSENSOR CONTROLBOX INDICATOR
LIGHT
Table 2. Planmeca ProSensor ControlBox indicator light explanation
NOTE The exposure can only be taken when the Planmeca
ProSensor ControlBox indicator light is green and
steady, not when the indicator light is flashing.
CONTROLBOX INDICATOR LIGHT
PLANMECA PROSENSOR STATUS
Off
Planmeca ProSensor system power off
Dim blue
Planmeca ProSensor system is off (not in
intraoral exposure-mode and the cable is
connected to the ControlBox)
Bright blue
Planmeca ProSensor system is on (Imaging
program communicates with the Planmeca
ProSensor system)
Slowly flashing blue
Waiting for Ready
Steady green
Waiting for Exposure
Rapidly flashing green
The exposure is taken and image is
transferred from the sensor to the ControlBox
Steady red
Error mode
Slowly flashing yellow
Service mode
Uploading ControlBox software
Slowly flashing blue, turns to slowly flashing
dim blue, then to quickly
flashing dim blue
Reading calibration
files from the sensor.
Flashing violet
ControlBox startup with back-up software