Planmeca ProSensor 15
CLEANING
User’s Manual
11
CLEANING
NOTE Before cleaning the system, always check that the X-
ray unit and the Planmeca ProSensor system are off
(Planmeca ProSensor ControlBox indicator light is
off).
11.1
Surfaces
The surfaces can be cleaned with a soft cloth damped in a
mild cleaning solution.
Stronger cleaning agents can be used for disinfecting the
surfaces. We recommend Dürr System-hygiene FD 333
or respective disinfecting solution.
11.2
Sensors and cables
Planmeca ProSensor sensors allow enhanced infection
control in the surgery.
As the sensor casing is hermetically sealed the sensors
can be immersed in disinfectant solution.
NOTE Always use appropriate instruments for cleaning the
sensors.
NOTE It is mandatory to carefully follow the disinfecting and
cleaning recommendations in order to not damage
the sensors.
CAUTION The sensors cannot be sterilized in
autoclave or UV oven.
Wipe up the sensor surface with a soft cloth damped into
a disinfectant solution. The sensors can be soaked in a
disinfection solution as long as there are no nicks in the
cable.
The recommendable disinfectant solutions are Dürr
System Hygiene FD 322 or FD 333 or similar product. The
immersion time with the Dürr disinfectants is 2 minutes.
If more effective disinfection or cold sterilization is
preferred for cleaning, we recommend the
Johnson&Johnson Cidex Opa high level disinfectant at a
minimum temperature of 20° C with maximum immersion
time of 8 minutes for a reuse period not to exceed 14
days.
NOTE Follow carefully the manufacturer’s
recommendations on immersion time and
recommended disinfectant liquids.