- 31 -
3.1 Center Ball
NOTE: Center ball will be above the midsagittal bar of the ball phantom. (Figure 43)
Figure 43
Click the second icon now and measure the top to bottom and the left to right.
Both should equal 7. (Figure 44)
Figure 44
The ball should be 7mm top to bottom and left to right. (Figure 45)
Figure 45