- 13 -
1.2 Left / Right
If the beam is too left or too right, loosen up the two screws at the right and
left positions on the collimator with a 2.5 mm allen (Figure 14)
Figure 14
Take another exposure until the beam is aligned.
1.3 Angled
If the beam is angled or not vertical, loosen up the four screws on the
collimator pointing up, down, left or right with a 3 mm allen. (Figure 15a &
15b)
Figure 15a
Figure 15b
Tighten up the left and right screws then take another exposure to check
alignment. If aligned then re-tighten the other two screws. If not, loosen the
two screws and move again.
Move onto Beam Check.