![Planer Kryo 1060 Series Operator'S Manual Download Page 37](http://html1.mh-extra.com/html/planer/kryo-1060-series/kryo-1060-series_operators-manual_1562042037.webp)
Kryo1060-180 Freezer
Note
These specifications are subject to change without notice.
Dimensions
External Internal
Height 112 cm
64 cm
Width
86 cm
50 cm
Depth
116 cm
50 cm
Weight
211kg (shipping weight including packaging) approx.
Capacity
4000 x 2ml ampoules
Circulation
Horizontal laminar flow.
Temperature
range
+40.0ºC to -100ºC.
Warning! The freezer is fitted with a manually-resetable
thermal cutout to prevent over heating. This will operate if
the chamber is programmed to run above +40.0ºC and will
require a Service Engineer to reset it.
Cooling
medium
Liquid nitrogen 22 - 30 psi.
Heater
1000W.
Sensors
Control and sample: 4-wire Platinum resistance
thermometer. Sensors are linearised in software to
international standards that utilise a 4096-point lookup
table based on BS EN 60751:1996. Calibration facility
provided.
Accuracy
±0.5ºC at a hold at 0ºC (dynamic accuracy depends on
actual program, e.g. rate of change of temperature).
Heating rates 0.01ºC/min to 1ºC/min.
Cooling rates -0.01ºC/min to -5ºC/min.
Kryo1060 Specifications