Figure 3.19
1
If your CIU-003 functions properly, COM-port cable is connected correctly, appropriate COM
port is selected, and the firmware to be downloaded is compatible with the Analyzer, the process of
downloading will start automatically. When the downloading is over, a pop-up window will
announce that the task has been successfully completed. The Analyzer will reboot and turn on in
the similar mode as if first time switched on.
CAUTION
Do not interrupt the loading process. This can lead to malfunction of
the Analyzer. But, if this does occur, repeat the process of program update.
4. MAINTENANCE
Required maintenance is limited to following the instructions related to proper operation, storage
and shipment, which are given in this Manual and also minor defects correction.
After the warranty period has expired and annually since then you should perform preventive
inspections covering check of controls, and reliability of the assembly.
5. TROUBLESHOOTING
Defect detection
: CIU-003 Analyzer does not switch on.
Possible reason
: One or both fuses are missing or damaged.
Methods of correction
: Insert two proper fuses into the fuse-holder.
Defect detection
: CIU-003 switches on, but cannot be configured by PC (see Section 3.3.2) or
connected to
ViewRSA
.
Possible reason
: Firmware hang-up.
Methods of correction
: Restart the Analyzer.
Possible reason
: Firmware failure.
20