Installation and Commissioning
•
The pressure tapping points should be prepared in accordance with the indications given for
the sockets. For more details, see e.g. rule VDE/VDI 3512, sheet3.
•
Suitable for sealing are sealing washers to DIN 16258.
•
The correct tightening torque is depending on material and shape of the used seal. It should
not exeed 80 Nm.
•
The mounting position should not be subject to strong vibration and radiation heat.
•
The mounting position which the transmitter is adjusted for, is indicated on the rating plate.
•
If the device is installed in a different position, the zero point may be offset. In this case, the
zero point should be readjusted as described here under.
•
The transmitters are immediatly ready for service after the pressure and electrical connecti-
ons have been made.
Pressure-connection variants
Connection G ½ B male
Connection G ¼ female
Socket DIN 16288
Connector
PUM04, PUM06 Instruction manual 12/2017
page 3
*
*
*
* 101 mm at thin-film technic (from range R79: 0...40 bar)
** 65 mm at thin-film technic (from range R79: 0...40 bar)
**
**