Programming lock
In order to prevent unauthorised adjustment of the device, a programming lock can be set. If the
menu item "PrG" is set to "Loc" in the extended menu, the programming lock is set. All values can
still be read but can't be edited. When trying to edit a value by means of the arrow keys, "Loc" is
displayed as long as the key is pressed. The functions "reS" and Rs.HL" are locked as well.
Error messages
If an error is detected, a corresponding error message appears that must be acknowledged by press-
ing any key.
Possible error messages:
E.10
A data error was detected in the saved settings. This could be due to strong electro-
magnetic interference or a component fault.
Action:
Press (E) and confirm "RES" by pressing "Yes". The factory settings will be restored
for all adjustable parameters and all minimum and maximum values will be deleted.
Enter the data again from the beginning.
E.12
An error was detected in the saved calibration data. This could be due to strong elec-
tromagnetic interference or a component fault.
Action:
Disconnect then reconnect the supply voltage to the instrument. If the error persists,
the instrument must be returned to the factory for recalibration or repair.
E.21
A communication error was detected within the instrument. This could be due to
strong electromagnetic interference or a component fault.
Action:
Press (E). If the error persists, disconnect then reconnect the supply voltage to the in-
strument. If the error still persists, please contact our service department.