PKP DU06 Series Instruction Manual Download Page 9

Mathematical parameters <Math.Parms> Menu 3

Characteristic diagram K-Factor – K-Factor

This function is used for adaption of the characteristic diagram up to formula 

Qkorr = Qist * K;

         Qist     -measured flow;
         Qkorr  -corrected flow;
         K         - absolute term (multiplicative)

It is used if flow conditions have changed referring to the original calibration

Advice:

This function is protected due to a separate Code (KO_Code). If the code is required
please ask
For it at the manufacturer

Characteristic diagram dynamical offset – Dynamical Offset

This function is used for adaption of the characteristic diagram up to formula:

Qkorr = Qist + DynOffset*(Qist/QmaxPhys);

With:  Qist              -measured flow;
           Qkorr           -corrected flow;
           QmaxPhys   - max. flow in pipe at v=5m/sec
           DynOffset   - absolute term (additive)
   

Advice:

This function is protected due to a separate Code (KO_Code). If the code is required
please ask for it at the manufacturer

Integration Time

The DU06 needs for one flow measurement approx. 80 ms. It calculates the indicated flow
out of a minimum of 10 und max. 100  measures. This means an adjustment range of the

Integration Time

  of 0,8 sec until 8 sec. The average value is detected by the method of

moving exponential average value. 

DU06 Instruction manual 09/2019

page  9

Summary of Contents for DU06 Series

Page 1: ...Instruction Manual DU06 Ultrasonic flowmeter with display PKP Prozessmesstechnik GmbH Borsigstra e 24 D 65205 Wiesbaden Nordenstadt Tel 49 0 6122 7055 0 Fax 49 0 6122 7055 50 Email info pkp de...

Page 2: ...vent of damage due to improper use non observance of this operating manual use of insufficiently qualified personnel and unauthorized modification of the device Proper Usage The ultrasonic flow meter...

Page 3: ...on versions are available version made of brass or stainless steel flow max 80 l min 1 version made of stainless steel flow max 180 l min 2 version made of stainless steel flow max 1000 l min All vers...

Page 4: ...urrent voltage frequency output 4 pole plug included in delivery Pin 1 4 20mA 0 10V Freq Out Pin 2 GND Pin 3 GND Pin 4 24VDC 15 Relay outputs 6 pol plug included in delivery Pin 1 Relay 2 normally ope...

Page 5: ...parameter Output Parms Menu 1 Relay 1 setting switch point of relay 1 Relay 2 setting switch point of relay 2 If the actual measured flow exceeds falls below one of the adjusted switch points the re...

Page 6: ...erated Freq Out 0 Hz is generated Assignment of analog output signal If the analogue output is used the following assignment must be observed 20 mA corresponds to the Max Flow Limit 4 mA corresponds t...

Page 7: ...c Litre per second l min Litre per minute l h Litre per hour m min Cubic meter per minute m h Cubic meter per hour gal sec Gallons per second gal min Gallons per minute gal h Gallons per hour Percent...

Page 8: ...e flow passing the DU06 during power failure was not counted Parameter protection Pin Pin Code is for protecting parameter of the device Changing them without authority will be avoided If the Pin is a...

Page 9: ...fset This function is used for adaption of the characteristic diagram up to formula Qkorr Qist DynOffset Qist QmaxPhys With Qist measured flow Qkorr corrected flow QmaxPhys max flow in pipe at v 5m se...

Page 10: ...s and protected due to manufacturers code 4 20 mA K value of current output 4 20 mA K Value This function is for special applications and protected due to manufacturers code 0 10 V K value of voltage...

Page 11: ...dium The DU06 indicates no negative flow If there is 0 flow indication while there is flow flow direction has to be changed in the program Functional check It is the responsibility of the operator to...

Page 12: ...time difference is proportional to the flow rate In contrast to other types of ultrasonic flowmeters the sensors are oriented parallel to the flow stream This results in an extremely compact meter wh...

Page 13: ...Dimensions Order Code Order number DU06 Ultrasonic flowmeter and counter with local indication E V R025 0 0 Material version M brass housing sensors st steel with R male connection only E st steel ho...

Reviews: