Please mount them properly only at the special mounting points.
Immediately unplug your unit from the wall outlet and refer servicing to qualified
service personnel under the following conditions:
•
when the power supply cord or plug is damaged
•
if liquid has been spilled or objects have fallen into the unit
•
if the unit has been exposed to rain or water
•
if the unit has been dropped or the housing is damaged
6. instructions for checking the function of the system / quick
start
When you open the package, please check to make sure all the necessary parts are
complete (see packing list) and nothing is damaged.
Check the electronics and the actuator for any visible damage:
•
The top and bottom plate of the actuator (if it does not have another shape) should
be parallel each to each other, without scratches.
•
If there is any damage to the system please contact our local representative
immediately!
•
If the packaging material is damaged please confirm this with the shipping
company.
•
The power switch is OFF.
•
Check that the available main voltage is the same as the label on the amplifiers
back side.
•
Connect the power supply.
•
Connect the piezo-element by using the „OUT“ connector.
•
Connect the measurement system on the „SENSOR“ connector, if availably.
•
Switch on the system by using the POWER switch on the front panel.
•
The green “ON” LED shines permanently. It follows a short selftest of all parts.
•
Always, when the system is switched on, it slowly goes through the voltage range
from -20V to 130V. This is to protect the actuator against depolarisation effects.
•
Switch on the closed loop by pressing the closed loop feeler in. The “CL” LED will
shine.
•
Now you can set the position with the OFFSET potentiometer.
•
If the “OOR” LED lights up there is an error.
•
Please switch off the device.
•
The function check is complete.