Hardware:
The output for trigger signals is the 9-pin D-Sub connector
“ANALOG”.
The pin 7 (trigger out) and pin 8 (ground) are used.
sketch 4: internal circuit
8.7 scan function
The scan allows the output of exactly one period of the function generator (sine or triangle). This
will require that the relevant functional parameters be set (frequency, amplitude, offset, or
symmetry). The function generator does not need to be activated (gfkt,0). To select a scan, set the
scan type to sine (sct,1) or triangle (sct,2).
The scan starts with
“ss,1”. During a scan, the status can be requested with "ss<CR>". The answer
“ss,2” means the scan is still running, “ss,0” indicates the scan is complete. A running scan can be
aborted by setting the status to zero (ss,0<CR>).