Geschäftsführer: Dr. Bernt Götz • Telefon 03641/66880 • Fax 03641/668866 • www.piezosystem.com
51
monsrc
monitor output
(0 = default)
0 = position in closed loop
1 = command value
2 = controller output voltage
3 = closed loop deviation incl. sign
4 = absolute closed loop deviation
5 = actuator voltage
6 = position in open loop
cl
open loop / closed loop
0 = open loop
1 = closed loop
kp
proportional term
0 to 999.0
ki
integral term
0 to 999.0
kd
differential term
0 to 999.0
sstd
set default values
------
notchon
notch filter
0 = off
1 = on
notchf
notch filter frequency
0 to 20000 [Hz]
notchb
bandwidth (-3dB)
0 to 20000 (max. 2 * notch_fr) [Hz]
Ipon
low pass filter
0 = off
1 = on
Ipf
low pass cut frequency
1 to 20000 [Hz]
gfkt
internal function generator
(see table 12)
0 = off
1 = sine
2 = triangle
3 = rectangle
4 = noise
5 = sweep
gasin
generator amplitude sine
0 to 100 [%]
gosin
amplitude offset sine
0 to 100 [%]
gfsin
generator frequency sine
0.1 to 9999.9 [Hz]
gatri
generator amplitude triangle
0 to 100 [%]
gotri
amplitude offset triangle
0 to 100 [%]
gftri
generator frequency triangle
0.1 to 9999.9 [Hz]
gstri
symmetry of triangle
0.1 to 99.9 [%]
default = 50 %
garec
generator amplitude rect.
0 to 100 [%]
gorec
amplitude offset rectangle
0 to 100 [%]
gfrec
generator frequency rectangle
0.1 to 9999.9 [Hz]
gsrec
symmetry of rectangle
0.1 to 99.9 [%]
default = 50 %
ganoi
generator amplitude noise
0 to 100 [%]
gonoi
amplitude offset noise
0 to 100 [%]
gaswe
generator amplitude sweep
0 to 100 [%]
goswe
amplitude offset sweep
0 to 100 [%]
gtswe
generator sweep time
0.4 to 800 [sec/decade]
sct
scan type
0 = scan function off
1 = sine scan
2 = triangle scan
ss
start scan
without value: request scan state
1 = starts scan