Geschäftsführer: Dr. Bernt Götz • Telefon 03641/66880 • Fax 03641/668866 • www.piezosystem.com
46
7.2.4. measurement module ECP1
width of the module itself
6TE = 30,5mm
connector for measurement
LEMO 0S.650
range
200µm or 500µm
sensitivity
100mV/µm or rather 40mV/µm
output
10V
power supply
15VDC /
50mA
resolution
0.005% FS
temperature range
sensor: -50 ... +150 °C
cable: -50 ... +150 °C
electronic: +10 ... +40 °C
humidity range
5 ... 95% (not condensing)
sensor diameter
10mm
weight ( cable)
56g
cable length
1.6m
min. diameter measure object
12mm
temperature stability electronic
0.005% FS / K
long time stability
0.04% FS / Month
european standards
EN 50081-1, EN 50082-2
protection standard (electronic
and sensor)
IP 40
linearity
typ. < 0.05 % FS
repeatability
typ. < 0.04 % FS
table 13: technical data measurement module ECP1
8.
operation
8.1. initiation
First check that your system uses the correct supply voltage (see the label on the
backside of the casing). Please connect the device with the wall outlet by using
the power cord. The piezo-actuators use operation voltages up to +150Volt.
Please be careful for any electric hazards. Do not remove the actuator plug until
the amplifier has been switched off. Piezoelectric actuators such stacks or various
tables work electrically as a capacitance. These elements are able to store
electrical energy over a long period (up to days) and the stored energy may be
dangerous.
When using the ENV 40 CSG/CCP please be sure to notice that before switching
on the system, that the closed-
loop switch is set to the “off” position.
Turn all potentiometers to maximal counter clockwise position. Remove external
modulation signals from MOD plug. Make sure all cable from the actuators (OUT
and SENSOR) are plugged into the correct amplifier.