background image

Accessories

Included:        

 

 

 

 

                 Part Number

  Four “AA” Alkaline batteries, Certificate of Calibration

  Evolution RTD Wire Kit   

            

 

                      020-0208

    2 Red & 2 Black Leads with Retractable Shield Banana Plugs & Spade Lugs

Optional:       

 

 

 

 

 

    Part Number

  Rubber Boot       

 

 

       

    

                      020-0209

  Small Carrying Case with PIE Logo (fits unit with or without boot)               020-0205

  Ni-MH 1 Hour Charger with 4 Ni-MH AA Batteries        

                      020-0103

    (100-120 V AC input for North America Only)

Ordering Information

PIECAL 510B 

(Choose one of the ordering codes below)

:

PIECAL 510B-Pt-100-1 (

a

=1.3850)

PIECAL 510B-Pt-100-2 (

a

=1.3902)

PIECAL 510B-Pt-100-3 (

a

=1.3916)

PIECAL 510B-Pt-100-4 (

a

=1.3926)

PIECAL 510B-Pt-1000-5 (

a

=1.3850)

PIECAL 510B-Pt-1000-6 (

a

=1.375)

PIECAL 510B-Cu-10
PIECAL 510B-Cu-50
PIECAL 510B-Ni-110
PIECAL 510B-Ni-120

PIECAL 511B: 

Includes all ten RTD Types listed above

More Than a Simple Boot

The optional boot provides 
more than just protection. Flip 
out the tilt stand and free up 
both hands for calibration 
adjustments.

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

TestEquipmentDepot.com 

Summary of Contents for PIECAL 510B

Page 1: ...atinum 100 alpha 3850 3902 3926 1000 alpha 3850 3750 Ohm Copper 10 50 Ohm Nickel 100 and 120 Ohm Easily set any value quickly to within 0 1 with the adjustable digital potentiometer DIAL plus store any three temperatures for instant recall with the EZ CHECK switch Compatible with all process instruments Connect directly to the RTD inputs of smart transmitters PLCs DCS and multichannel recorders an...

Page 2: ...ANGING BATTERIES Low battery is indicated by BAT on the display Approximately one to four hours of typical operation remain before the PIECAL 510B 511B will automatically turn off To change the batteries remove the optional rubber boot remove the battery door from the back of the unit by sliding the door downward This allows access to the battery compartment Replace with four 4 AA 1 5V batteries b...

Page 3: ...Menu will automatically exit after 15 seconds of inactivity countdown timer is displayed TEMP UNITS pressing the knob will toggle between C and F RTD 510 pressing the knob will toggle between the factory configured RTD Pt Cu Ni and Ohms 511 pressing the knob will cycle through the ten different RTD types Pt Cu Ni at different base resistances and alpha values and Ohms AUTO OFF If AUTO OFF is ON th...

Page 4: ...high SPAN output temperature by moving the EZ CHECK switch to the HI position and turn the e EZ Dial knob until the desired temperature is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 2 Store your low ZERO output temperature by moving the EZ CHECK switch to the LO position and turn the e EZ Dial knob until the desired temperature i...

Page 5: ... 100 0 to 700 0 700 0 to 850 0 0 2 0 3 0 4 328 0 to 212 0 212 0 to 1292 0 1292 0 to 1562 0 0 4 0 6 0 7 Pt 1000 Ohm DIN IEC JIS 1989 1 3850 0 00385 200 0 to 200 0 200 0 to 600 0 600 0 to 850 0 0 2 0 3 0 4 328 0 to 392 0 392 0 to 1112 0 1112 0 to 1562 0 0 4 0 6 0 7 Pt 1000 Ohm Hy Cal HVAC 1 3750 0 00375 200 0 to 200 0 200 0 to 600 0 600 0 to 850 0 0 2 0 3 0 4 328 0 to 392 0 392 0 to 1112 0 1112 0 to...

Page 6: ...ration left Protection against misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas PIECAL 510B 511B Specifications Unless otherwise indicated all specifications are rated from a nominal 23 C 70 RH for 1 year from calibration Source Accuracy From 1 to 10 2 mA External Excitation Current ...

Page 7: ... given under our warranty No responsibility is accepted for damage loss or other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Information This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration Test Data is availa...

Page 8: ...formation PIECAL 510B Choose one of the ordering codes below PIECAL 510B Pt 100 1 a 1 3850 PIECAL 510B Pt 100 2 a 1 3902 PIECAL 510B Pt 100 3 a 1 3916 PIECAL 510B Pt 100 4 a 1 3926 PIECAL 510B Pt 1000 5 a 1 3850 PIECAL 510B Pt 1000 6 a 1 375 PIECAL 510B Cu 10 PIECAL 510B Cu 50 PIECAL 510B Ni 110 PIECAL 510B Ni 120 PIECAL 511B Includes all ten RTD Types listed above More Than a Simple Boot The opti...