background image

q

 

EZ-CHECK™ SWITCH

SOURCE:

  Instantly  output  two  preset  RTD 

temperatures by moving the EZ-CHECK™ switch to 
the 

“LO” 

position  or 

“HI”

  position.  For  fast  three 

point  checks  select  the 

“DIAL”

  position.  The 

PIECAL  510B/511B  will  remember  the  last 

“DIAL” 

value, even with the power off.

These  values  can  easily  be  changed  to  suit  the 
calibration requirements.

w

 

SOURCE/OFF Switch

Select “

SOURCE

” to output in °C, °F or ohms.  

e

 

EZ-DIAL™ KNOB

SOURCE:

  Turn  the  knob  to  adjust  the  output 

level.  Turn  clockwise  to  increase  the  output, 
counter clockwise to decrease the output in 0.1°  
steps  at  a  time.  Push  down  and  turn  the 
EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store 
desired EZ-Check™ HI/LO points in SIMULATE 
mode. 

Double  click  the  knob  to  get  into  the  PIECAL 
510B/511B 

Configuration 

Mode. 

Use 

configuration to  select °C or °F, RTD Type (511B 
only) and Auto Off On/Off.

CHANGING BATTERIES

Low battery is indicated by 

“BAT”

 on the display. 

Approximately  one  to  four  hours  of  typical 
operation remain before the PIECAL 510B/511B 
will automatically turn off. To change the batteries; 
remove  the  optional  rubber  boot,  remove  the 
battery door from the back of the unit by sliding 
the  door  downward.  This  allows  access  to  the 
battery  compartment.  Replace  with  four  (4)  “AA” 
1.5V batteries being careful to check the polarity. 
Replace the battery door and replace the boot. All 
stored  configuration  options  (RTD  Type, 
EZ-CHECK  Memories, etc., are reset to factory 
settings when the batteries are removed.

Note:

  Alkaline  batteries  are  supplied  and 

recommended  for  maximum  battery  life  and 
performance.

Basic Operation

3 2 2 - 1

THERMOCOUPLE

CALIBRATOR

J, T, E, K, mV

HI

Pt100 

α

=385

AUTOMATED

RTD

CALIBRATOR

SOURCE

°C

3W   2W      1W  4W

     

          

    

OFF

Push & Hold

to Store

Push & Turn

for

Fast Dialing

Double Click

for Options

SOURCE

HI

SET

LO

MULTIPLE RTD TYPES

RTD SIMULATOR

100.0

(Optional Boot Shown)

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

TestEquipmentDepot.com 

Summary of Contents for PIECAL 510B

Page 1: ...atinum 100 alpha 3850 3902 3926 1000 alpha 3850 3750 Ohm Copper 10 50 Ohm Nickel 100 and 120 Ohm Easily set any value quickly to within 0 1 with the adjustable digital potentiometer DIAL plus store any three temperatures for instant recall with the EZ CHECK switch Compatible with all process instruments Connect directly to the RTD inputs of smart transmitters PLCs DCS and multichannel recorders an...

Page 2: ...ANGING BATTERIES Low battery is indicated by BAT on the display Approximately one to four hours of typical operation remain before the PIECAL 510B 511B will automatically turn off To change the batteries remove the optional rubber boot remove the battery door from the back of the unit by sliding the door downward This allows access to the battery compartment Replace with four 4 AA 1 5V batteries b...

Page 3: ...Menu will automatically exit after 15 seconds of inactivity countdown timer is displayed TEMP UNITS pressing the knob will toggle between C and F RTD 510 pressing the knob will toggle between the factory configured RTD Pt Cu Ni and Ohms 511 pressing the knob will cycle through the ten different RTD types Pt Cu Ni at different base resistances and alpha values and Ohms AUTO OFF If AUTO OFF is ON th...

Page 4: ...high SPAN output temperature by moving the EZ CHECK switch to the HI position and turn the e EZ Dial knob until the desired temperature is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 2 Store your low ZERO output temperature by moving the EZ CHECK switch to the LO position and turn the e EZ Dial knob until the desired temperature i...

Page 5: ... 100 0 to 700 0 700 0 to 850 0 0 2 0 3 0 4 328 0 to 212 0 212 0 to 1292 0 1292 0 to 1562 0 0 4 0 6 0 7 Pt 1000 Ohm DIN IEC JIS 1989 1 3850 0 00385 200 0 to 200 0 200 0 to 600 0 600 0 to 850 0 0 2 0 3 0 4 328 0 to 392 0 392 0 to 1112 0 1112 0 to 1562 0 0 4 0 6 0 7 Pt 1000 Ohm Hy Cal HVAC 1 3750 0 00375 200 0 to 200 0 200 0 to 600 0 600 0 to 850 0 0 2 0 3 0 4 328 0 to 392 0 392 0 to 1112 0 1112 0 to...

Page 6: ...ration left Protection against misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas PIECAL 510B 511B Specifications Unless otherwise indicated all specifications are rated from a nominal 23 C 70 RH for 1 year from calibration Source Accuracy From 1 to 10 2 mA External Excitation Current ...

Page 7: ... given under our warranty No responsibility is accepted for damage loss or other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Information This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration Test Data is availa...

Page 8: ...formation PIECAL 510B Choose one of the ordering codes below PIECAL 510B Pt 100 1 a 1 3850 PIECAL 510B Pt 100 2 a 1 3902 PIECAL 510B Pt 100 3 a 1 3916 PIECAL 510B Pt 100 4 a 1 3926 PIECAL 510B Pt 1000 5 a 1 3850 PIECAL 510B Pt 1000 6 a 1 375 PIECAL 510B Cu 10 PIECAL 510B Cu 50 PIECAL 510B Ni 110 PIECAL 510B Ni 120 PIECAL 511B Includes all ten RTD Types listed above More Than a Simple Boot The opti...