Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Configuration
Double Click Menus - Main Page
Move
w
POWER SWITCH to
“SOURCE” or
“READ”.
Turn the
e
DIAL KNOB to move through the two
pages of menus. Shown are the MAIN menus for
each function. Turn the
e
knob to scroll thru the
menus and press the
e
knob to select.
Available choices are shown in grey.
Source mA & Simulate
2 Wire Transmitters
Read mA, Power/Measure Transmitters
& Leak Detect
Source V
Read V & mV
Source pH
Source & Read Frequency
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
V#.##
Double Click Menus - Main Page
Source & Read Thermocouples
Source RTD
Read RTD
Source Ohms
Read Ohms
Read Pressure
MAIN
> EXIT
FUNCTION
mA
MODE
SOURCE
2W SIM
UNITS
mA
%
HART 250
Ω
ON
OFF
> EXIT
FUNCTION
mA
MODE
READ
PWR MEAS
UNITS
mA
%
HART 250
Ω
ON
OFF
> EXIT
FUNCTION V
RANGE 10V
100mV 1V
> EXIT
FUNCTION V
RANGE 10V
1V 60V 100mV
> EXIT
FUNCTION pH
> EXIT
FUNCTION FREQ
RANGE 20KHZ
10000HZ 1000HZ 2000CPM
MAIN
> EXIT
FUNCTION T/C
UNITS
°C
°F
T/C
TYPE
J
K E T R S B N L U G C D P
COLD JUNC ON
OFF
>EXIT
FUNCTION RTD
UNITS °C
°F
RTD Pt 100 a=3850 [*RTD Types - See Read RTD]
>EXIT
FUNCTION RTD
UNITS °C
°F
RTD Pt 100
α
=3850,
α
=3902,
α
=3916,
α
=3926
Pt 1000
α
=3850;
Cu
10
α
=4274, Cu 50
α
=4280
Ni
120
α
=6720, Ni 110
α
=5801
> EXIT
FUNCTION OHMS
RANGE 400Ω
4000Ω
> EXIT
FUNCTION OHMS
RANGE 400Ω
4000Ω CONT
> EXIT
FUNCTION PRESSURE
UNITS psi
inH2O* ftH2O* mmH2O*
cmH2O* mH2O* inHg mHg cmHg mmHg
torr kg/cm2 kg/m2 hPa kPa MPa bar mbar
atm oz/in2 lb/ft2
*Engineering unit available at 4°C, 20°C and
60°F.