Configuration
Double Click Menus - MAIN Page
Move
w
POWER SWITCH to
“SOURCE”
or
“READ”
.
Double click the EZ-DIAL knob to access the
Double Click Menus. Shown are the
MAIN
menus for each function. Turn the knob to scroll
thru the menus and press the knob to select.
Available choices are shown in grey.
Source RTDs
Read RTDs
Source & Read Ohms
Source mA &
Simulate 2 Wire Transmitters
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
V#.##
MAIN
> EXIT (1/3)
FUNCTION OHMS
RTD mA
RANGE
400 Ω
4000 Ω
MAIN
> EXIT (1/2)
FUNCTION
mA
MODE
SOURCE
2W SIM
UNITS
mA
%
HART 250Ω
ON
OFF
MAIN
>EXIT (1/3)
FUNCTION RTD
UNITS
°C
°F
RTD Pt 100
a
=3850
[*See Read RTD]
MAIN
>EXIT (1/3)
FUNCTION RTD
UNITS
°C
°F
RTD Pt 100
a
=3850,
Pt 200
a
=3850,
Pt 500
a
=3850, Pt 1000
a
=3850,
Pt 100
a
=3902, Pt 100
a
=3916,
Pt 100
a
=3926,
Cu 10
a
=4274, Cu 50
a
=4280,
Ni 120
a
=6720
Pt 10
a
=3850, Pt 50
a
=3850
Read mA, Power/Measure
Transmitters & Leak Detect
Turn the
e
DIAL KNOB to move through the two
to four pages of menus. Press the
e
DIAL
KNOB to toggle between
OFF
and
ON
or to
scroll through the settings.
MAIN
> EXIT (1/3)
FUNCTION
mA
MODE
READ
PWR MEAS
UNITS
mA
%
HART 250Ω
ON
OFF
Double Click Menu - RTD DISPLAY
DISPLAY OHMS
- If DISPLAY OHMS is ON the
Ohms value corresponding to the sourced or
measured temperature is displayed.
SENSOR mA
- If SENSOR mA is ON the
excitation current supplied from the RTD
measuring instrument is displayed (SOURCE
only).
Double Click Menu - OHMS DISPLAY
SENSOR mA
- If SENSOR mA is ON the
excitation current supplied from the resistance
measuring instrument is displayed (SOURCE
only).
RTD DISPLAY
(SOURCE)
> EXIT (2/4)
DISPLAY OHMS OFF
ON
SENSOR mA* OFF
ON
OHMS DISPLAY
> EXIT (2/4)
SENSOR mA* OFF
ON
RTD DISPLAY
(READ)
> EXIT (2/3)
DISPLAY OHMS OFF
ON