Configuration

Double Click Menus - MAIN Page

Move 

POWER SWITCH to 

“SOURCE”

 or 

“READ”

Double click the EZ-DIAL knob to access the 
Double Click Menus. Shown are the 

MAIN

 

menus for each function. Turn the knob to scroll 
thru the menus and press the knob to select. 
Available choices are shown in grey.

Source RTDs

Read RTDs

Source & Read Ohms

Source mA &

Simulate 2 Wire Transmitters

DOUBLE CLICK

EZ-DIAL KNOB

FOR CONFIGURATION

    V#.##

           MAIN

>  EXIT (1/3)   

 

 

      

 

FUNCTION    OHMS 

RTD mA

 

RANGE   

 400 Ω 

4000 Ω

           MAIN

>  EXIT  (1/2)   

 

      

 

FUNCTION   

      mA

 MODE   

 

SOURCE 

2W SIM

 

UNITS 

 

 

      mA 

%

 

HART 250Ω   

      ON 

OFF

             

MAIN

>EXIT (1/3)   

 

 

      

  FUNCTION RTD

 

UNITS 

  °C 

°F

 

RTD    Pt 100 

a

=3850  

[*See Read RTD]

          MAIN

   

>EXIT (1/3)   

 

 

      

  FUNCTION RTD

 

UNITS 

    °C 

°F

 

RTD       Pt 100 

a

=3850,

 

Pt 200 

a

=3850,    

 

Pt 500 

a

=3850, Pt 1000 

a

=3850,

 

Pt 100 

a

=3902, Pt 100 

a

=3916, 

 

Pt 100 

a

=3926,

 

Cu 10 

a

=4274, Cu 50 

a

=4280,

 

Ni 120 

a

=6720

 

Pt 10 

a

=3850, Pt 50 

a

=3850

Read mA, Power/Measure 

Transmitters & Leak Detect 

Turn the 

e

 DIAL KNOB to move through the two 

to four pages of menus. Press the 

e

 DIAL 

KNOB to toggle between 

OFF

 and 

ON

 or to 

scroll through the settings. 

 

             MAIN

>  EXIT  (1/3)   

      

 

FUNCTION   

      mA

 MODE   

 

READ 

PWR MEAS

 

UNITS 

 

 

      mA 

%

 

HART 250Ω   

      ON 

OFF

Double Click Menu - RTD DISPLAY

DISPLAY OHMS

 - If DISPLAY OHMS is ON the 

Ohms value corresponding to the sourced or 
measured temperature is displayed.

SENSOR mA

 - If SENSOR mA is ON the 

excitation current supplied from the RTD 
measuring instrument is displayed (SOURCE 
only).

Double Click Menu - OHMS DISPLAY

SENSOR mA

 - If SENSOR mA is ON the 

excitation current supplied from the resistance 
measuring instrument is displayed (SOURCE 
only).

     RTD DISPLAY 

(SOURCE)

>  EXIT (2/4)   

 

 

      

 

DISPLAY OHMS    OFF 

ON

 

SENSOR mA*        OFF 

ON

     OHMS DISPLAY

>  EXIT (2/4)   

 

 

      

 

SENSOR mA*        OFF 

ON

     RTD DISPLAY 

(READ)

>  EXIT (2/3)   

 

 

      

 

DISPLAY OHMS    OFF 

ON

Summary of Contents for 311Plus

Page 1: ...for simple hookups Carry it without worry it comes protected with a rubber boot and rugged low profile switches Easy to operate even in dark areas with the backlit display Calibrate directly in temper...

Page 2: ...ectronics 82 East Main Street Suite 3 14 Webster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Magnetic Hanging Strap Optional Deluxe Hands Free Carrying Case Included Evo...

Page 3: ...and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for three more seconds to...

Page 4: ...0 a 3850 Pt 200 a 3850 Pt 500 a 3850 Pt 1000 a 3850 Pt 100 a 3902 Pt 100 a 3916 Pt 100 a 3926 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Pt 10 a 3850 Pt 50 a 3850 Read mA Power Measure Transmitters Leak...

Page 5: ...EXIT 4 4 AUTO OFF ON OFF BACKLIGHT ON OFF STEPS RAMP 2 3 5 11 RAMP STEP RAMP TIME 5 6 7 8 9 10 15 20 25 30 60 Double Click Menu FEATURES Double Click Menu mA DISPLAY MODE pressing the knob will cycle...

Page 6: ...he EZ CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the...

Page 7: ...nections are made at the top of the calibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated Connections REA...

Page 8: ...made to a 4 wire sensor and the 311Plus indicates that only Wires 1 2 4 are connected There may be a loose connection or a break in wire 3 somewhere between the sensor and the 311Plus Only a calibrato...

Page 9: ...calibrated and connect the PIE 311Plus to the inputs of the device using 2 3 or 4 wires The sensor current generated by the instrument will be indicated on the display followed by the word FIXED or PU...

Page 10: ...Detect Transmitter Supply C 3 2 1 4 100 0 OFF OVERLOAD Double Click Menu HI SET LO SOURCE MAX READ MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4...

Page 11: ...AD MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4 Wire Detection Loop Leak Detect Transmitter Supply mV 40 000 All RTD Ohms Connections Voltage Ou...

Page 12: ...such as mid range using the SET position on the EZ CHECK switch The output is adjusted in 0 001 mA 0 01 increments by turning the knob e Press and turn the knob for faster dialing with 0 100 mA 1 00 i...

Page 13: ...and disconnect the PIE 311Plus Signals above 24 mA are current limited by protection circuitry with OVERRANGE flashed on the display and the red OVERLOAD LED lit To Sensor Typical 2 Wire Transmitter D...

Page 14: ...oose READ if your would like the mA display in percent of 4 20 milliamps PWRM is POWER MEASURE which uses the internal loop supply of the 311Plus to power up the transmitter while indicating the curre...

Page 15: ...ster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Operating Instructions Typical 2 Wire Transmitter IN REF OUT 311 2 F HART 5 6 PWRM 12 523 mA Pt 100 3850 00 51 mA LEAK E...

Page 16: ...PIE 311Plus supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop The current passed by the transmitter will be accurately displayed by the 311Plus along with an indication of leak...

Page 17: ...inst misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Ohms RTD 3 Wire 4 Wire Ac...

Page 18: ...003 mA Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nominal Source mA Power Measure Two Wire Transmitters PWRM LEAK Ranges and Resolution 0 000 to 24 000 mA or 25 00 to 125 00 of...

Page 19: ...0 842 0 to 1256 0 1256 0 to 1562 0 0 1 0 2 0 3 0 4 0 5 Pt 500 Ohm DIN IEC JIS 1989 ITS 90 1 3850 0 00385 200 0 to 90 0 120 0 to 180 0 180 0 to 450 0 450 0 to 680 0 680 0 to 850 0 0 1 0 1 0 2 0 2 0 3 3...

Page 20: ...other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Inform...

Reviews: