PIE 311Plus Operating Instructions Manual Download Page 16

Page 16

Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Operating Instructions

mA OUT, % OUT (Percent of 4 to 20 mA)

Find current leaks in loops caused by ground faults, moisture or corrosion. The 311Plus 
simultaneously supplies power to a 2 Wire Transmitter (or loop with a transmitter) while displaying 
the 4 to 20 mA output and the amount of current leaking in the loop.

1)  Move the power switch 

to READ then Double Click the EZ-DIAL knob to get into the Menu. 

Turn the knob 

e

 to scroll through the settings and press the knob to make your selection. Select 

mA

 for the 

FUNCTION

 and 

PWR MEAS

 for the 

MODE

. Choose either 

mA

 or 

%

.

2)  Turn the knob 

e

 until the menu to the right appears.

3)  Turn the knob 

e

 to scroll through the settings and press the 

knob to make your selection. Turn on the 

LEAK DETECT

.

4) Connect the red source lead from the mA (+) jack of the 

311Plus to the plus (+) input of the device and the black source lead from the mA (-) to the minus 
(-).

The PIE 311Plus supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop. The 
current passed by the transmitter will be accurately displayed by the 311Plus along with an indication 
of leakage current at the top of the display. If there is an uncontrolled loop caused by a transmitter 
with upscale burnout and a sensor that is burned out or missing, or a short in the loop wiring, the 
311Plus will display "OVER RANGE”

Using Ground Leak Detection

         

mA DIAGNOSTIC

>  EXIT (2/3)        
 

LEAK DETECT

     OFF 

ON

Typical Error Conditions

The PIE 311Plus is supplying the loop voltage. 
A calibrated transmitter is limiting the loop 
current to 12.00 mA. An additional 0.51 mA is 
not controlled by the transmitter and is leaking 
somewhere in the loop.

The PIE 311Plus is supplying the loop voltage. 
There is a control loop error. This may be a 
transmitter (set for upscale burnout) with a 
bad or missing sensor, or a short in the loop. 
The red ERROR LED will also flash.

                  

PWRM

  

LEAK

          00.51  mA

 

    

12.506

 mA

5    6

HART

              

PWRM

     

    OVER RANGE

 

                           mA

5    6

HART

Note:

 Many loops with installed transmitters will 

normally indicate 0.01 to 0.02 mA leakage 
without significant control problem. Unstable 
readings may indicate loose connections or the 
presence of moisture.

All Milliamp Connections

Summary of Contents for 311Plus

Page 1: ...for simple hookups Carry it without worry it comes protected with a rubber boot and rugged low profile switches Easy to operate even in dark areas with the backlit display Calibrate directly in temper...

Page 2: ...ectronics 82 East Main Street Suite 3 14 Webster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Magnetic Hanging Strap Optional Deluxe Hands Free Carrying Case Included Evo...

Page 3: ...and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for three more seconds to...

Page 4: ...0 a 3850 Pt 200 a 3850 Pt 500 a 3850 Pt 1000 a 3850 Pt 100 a 3902 Pt 100 a 3916 Pt 100 a 3926 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Pt 10 a 3850 Pt 50 a 3850 Read mA Power Measure Transmitters Leak...

Page 5: ...EXIT 4 4 AUTO OFF ON OFF BACKLIGHT ON OFF STEPS RAMP 2 3 5 11 RAMP STEP RAMP TIME 5 6 7 8 9 10 15 20 25 30 60 Double Click Menu FEATURES Double Click Menu mA DISPLAY MODE pressing the knob will cycle...

Page 6: ...he EZ CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the...

Page 7: ...nections are made at the top of the calibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated Connections REA...

Page 8: ...made to a 4 wire sensor and the 311Plus indicates that only Wires 1 2 4 are connected There may be a loose connection or a break in wire 3 somewhere between the sensor and the 311Plus Only a calibrato...

Page 9: ...calibrated and connect the PIE 311Plus to the inputs of the device using 2 3 or 4 wires The sensor current generated by the instrument will be indicated on the display followed by the word FIXED or PU...

Page 10: ...Detect Transmitter Supply C 3 2 1 4 100 0 OFF OVERLOAD Double Click Menu HI SET LO SOURCE MAX READ MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4...

Page 11: ...AD MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4 Wire Detection Loop Leak Detect Transmitter Supply mV 40 000 All RTD Ohms Connections Voltage Ou...

Page 12: ...such as mid range using the SET position on the EZ CHECK switch The output is adjusted in 0 001 mA 0 01 increments by turning the knob e Press and turn the knob for faster dialing with 0 100 mA 1 00 i...

Page 13: ...and disconnect the PIE 311Plus Signals above 24 mA are current limited by protection circuitry with OVERRANGE flashed on the display and the red OVERLOAD LED lit To Sensor Typical 2 Wire Transmitter D...

Page 14: ...oose READ if your would like the mA display in percent of 4 20 milliamps PWRM is POWER MEASURE which uses the internal loop supply of the 311Plus to power up the transmitter while indicating the curre...

Page 15: ...ster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Operating Instructions Typical 2 Wire Transmitter IN REF OUT 311 2 F HART 5 6 PWRM 12 523 mA Pt 100 3850 00 51 mA LEAK E...

Page 16: ...PIE 311Plus supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop The current passed by the transmitter will be accurately displayed by the 311Plus along with an indication of leak...

Page 17: ...inst misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Ohms RTD 3 Wire 4 Wire Ac...

Page 18: ...003 mA Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nominal Source mA Power Measure Two Wire Transmitters PWRM LEAK Ranges and Resolution 0 000 to 24 000 mA or 25 00 to 125 00 of...

Page 19: ...0 842 0 to 1256 0 1256 0 to 1562 0 0 1 0 2 0 3 0 4 0 5 Pt 500 Ohm DIN IEC JIS 1989 ITS 90 1 3850 0 00385 200 0 to 90 0 120 0 to 180 0 180 0 to 450 0 450 0 to 680 0 680 0 to 850 0 0 1 0 1 0 2 0 2 0 3 3...

Page 20: ...other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Inform...

Reviews: