q

 

EZ-CHECK™ SWITCH

SOURCE:

 Instantly output two preset RTD 

temperatures or milliamp values by moving the 
EZ-CHECK™ switch to the 

“LO” 

position or 

“HI”

 

position.  For fast three point checks select the 

“SET”

 position. The PIE 311Plus will remember 

the last 

“SET” 

value, even with the power off.

These values can easily be changed to suit the 
calibration requirements. The values stored in the HI 
and LO positions are also used for Auto Stepping.

READ:

 Slide the switch to the SET position. The PIE 

311Plus will display the current temperature from the 
RTD sensor or the current milliamp value.  Slide the 
switch to HI and the highest value measured since 
turn-on or reset will be displayed;  slide the switch to  
LO and the lowest value measured since turn-on or 
reset will be displayed.

w

 

SOURCE/OFF/READ Switch

Select “

SOURCE

” to output in °C, °F, Ohms or 

milliamps.  
Select “

READ

” to read a RTD sensor, Ohms or 

milliamps. Select “

OFF

” to turn the unit off.

e

 

EZ-DIAL™ KNOB

SOURCE: Turn the knob to adjust the output 
level. Turn clockwise to increase the output, 
counter clockwise to decrease the output in 0.1°, 
0.01 Ω or 0.001 mA  steps at a time. Push down 
and turn the EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store 
desired EZ-Check™ HI/LO points in SOURCE 
mode. Continue to press and hold the knob for 
three more seconds to start the automatic 
stepping or ramping.

READ:  Press and hold the knob for two seconds 
to transfer the current measured value into the 
EZ-Check™ HI/LO points. This clears the HI/LO 
readings which will update as the measured 
value changes.

Double click the knob to get into the PIE 311Plus 
Configuration Mode. 

CHANGING BATTERIES

Low battery is indicated by a battery symbol on 
the display. Approximately one hour of typical 
operation remain after the first time the low 
battery symbol appears. To change the batteries; 
remove the rubber boot, remove the battery door 
from the back of the unit by sliding the door 
downward. This allows access to the battery 
compartment. Replace with four (4) “AA” 1.5V 
batteries being careful to check the polarity. 
Replace the battery door and the boot. All stored 
configuration options (T/C Type, EZ-CHECK 

 

Memories, etc.,) are reset to factory settings 
when the batteries are removed.

Note:

 Alkaline batteries are supplied and 

recommended.  Purchase the optional Ni-MH 
rechargeable batteries for maximum battery life.

Basic Operation

°C

3      2                1      4

     

           

    

100.0

OFF

OVERLOAD

Double Click 

Menu

HI

SET

LO

SOURCE

MAX

READ

MIN

READ

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Diagnostic RTD &

Milliamp Calibrator

2, 3, 4 Wire Detection

Loop Leak Detect ∙ Transmitter Supply

+

-

-

+

Summary of Contents for 311Plus

Page 1: ...for simple hookups Carry it without worry it comes protected with a rubber boot and rugged low profile switches Easy to operate even in dark areas with the backlit display Calibrate directly in temper...

Page 2: ...ectronics 82 East Main Street Suite 3 14 Webster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Magnetic Hanging Strap Optional Deluxe Hands Free Carrying Case Included Evo...

Page 3: ...and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for three more seconds to...

Page 4: ...0 a 3850 Pt 200 a 3850 Pt 500 a 3850 Pt 1000 a 3850 Pt 100 a 3902 Pt 100 a 3916 Pt 100 a 3926 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Pt 10 a 3850 Pt 50 a 3850 Read mA Power Measure Transmitters Leak...

Page 5: ...EXIT 4 4 AUTO OFF ON OFF BACKLIGHT ON OFF STEPS RAMP 2 3 5 11 RAMP STEP RAMP TIME 5 6 7 8 9 10 15 20 25 30 60 Double Click Menu FEATURES Double Click Menu mA DISPLAY MODE pressing the knob will cycle...

Page 6: ...he EZ CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the...

Page 7: ...nections are made at the top of the calibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated Connections REA...

Page 8: ...made to a 4 wire sensor and the 311Plus indicates that only Wires 1 2 4 are connected There may be a loose connection or a break in wire 3 somewhere between the sensor and the 311Plus Only a calibrato...

Page 9: ...calibrated and connect the PIE 311Plus to the inputs of the device using 2 3 or 4 wires The sensor current generated by the instrument will be indicated on the display followed by the word FIXED or PU...

Page 10: ...Detect Transmitter Supply C 3 2 1 4 100 0 OFF OVERLOAD Double Click Menu HI SET LO SOURCE MAX READ MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4...

Page 11: ...AD MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4 Wire Detection Loop Leak Detect Transmitter Supply mV 40 000 All RTD Ohms Connections Voltage Ou...

Page 12: ...such as mid range using the SET position on the EZ CHECK switch The output is adjusted in 0 001 mA 0 01 increments by turning the knob e Press and turn the knob for faster dialing with 0 100 mA 1 00 i...

Page 13: ...and disconnect the PIE 311Plus Signals above 24 mA are current limited by protection circuitry with OVERRANGE flashed on the display and the red OVERLOAD LED lit To Sensor Typical 2 Wire Transmitter D...

Page 14: ...oose READ if your would like the mA display in percent of 4 20 milliamps PWRM is POWER MEASURE which uses the internal loop supply of the 311Plus to power up the transmitter while indicating the curre...

Page 15: ...ster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Operating Instructions Typical 2 Wire Transmitter IN REF OUT 311 2 F HART 5 6 PWRM 12 523 mA Pt 100 3850 00 51 mA LEAK E...

Page 16: ...PIE 311Plus supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop The current passed by the transmitter will be accurately displayed by the 311Plus along with an indication of leak...

Page 17: ...inst misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Ohms RTD 3 Wire 4 Wire Ac...

Page 18: ...003 mA Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nominal Source mA Power Measure Two Wire Transmitters PWRM LEAK Ranges and Resolution 0 000 to 24 000 mA or 25 00 to 125 00 of...

Page 19: ...0 842 0 to 1256 0 1256 0 to 1562 0 0 1 0 2 0 3 0 4 0 5 Pt 500 Ohm DIN IEC JIS 1989 ITS 90 1 3850 0 00385 200 0 to 90 0 120 0 to 180 0 180 0 to 450 0 450 0 to 680 0 680 0 to 850 0 0 1 0 1 0 2 0 2 0 3 3...

Page 20: ...other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Inform...

Reviews: