Frequency compensation
Before taking any measurements using the probe, first check its
compensation and adjust it to match the channel inputs.
Connect the probe to a 2 V pk-pk, 1 kHz square wave source. Most PicoScope
oscilloscopes have a signal generator output marked GEN or AWG, which you
can configure to generate such a signal. Set the switch on the probe to the
X10 position. Adjust the trimmer until you see a flat-top square wave on the
display:
✓
CORRECT
✗
INCORRECT
✗
INCORRECT
Accessories and features
The probe is provided with several accessories designed to make probing
and measurement simpler. Please take a moment to familiarize yourself with
these accessories and their uses.
Accessories included
TA375
TA386
Retractable hook tip
1
1
Ground lead
1
1
Adjustment tool
1
1
Tip insulating sleeve
1
1
Cable marker
2
8
Optional accessories
Order code Description
TA384
Replacement rigid probe tips for TA375 and TA386, pack of 5
TA385
Replacement spring probe tips for TA375 and TA386, pack of 5