8
DO337-1
P2036 and P2056 User's Guide
Pico Technology
Adjustment procedure
The probe can be adjusted for low frequency (LF) compensation.
LF compensation is necessary whenever the probe is connected to a new oscilloscope input channel.
LF compensation matches the probe's series capacitance to the oscilloscope's input capacitance.
This matching assures good amplitude accuracy from DC to upper bandwidth limit frequencies. A
poorly compensated probe influences the overall system performance (probe + scope) and introduces
measurement errors resulting in inaccurate readings and distorted waveforms.
LF compensation is performed by connecting the probe to a suitable signal source such as:
• the CAL output on the oscilloscope front panel
• a good-quality 500 Hz square wave from the AWG output of a PicoScope
• a square wave output from an oscilloscope calibrator
and adjusting the LF compensation trimmer to obtain optimal square wave response. For examples
see figures below.
LF compensation trimmer
undercompensated
optimal
overcompensated