Specifications
This User's Guide supersedes all previously published material. Specifications that are not marked as
guaranteed are published as general information to the user. The instrument should have warmed up
for at least 20 minutes and the environmental conditions must not exceed the specified limits of the
probe. Note that specifications are subject to change without notice.
Electrical specifications
P2036
P2056
Attenuation ratio
*
10:1
(±2% at DC)
Voltage coefficient
0.00025 %/V
(at DC)
Probe bandwidth
300 MHz
500 MHz
(–3 dB)
Probe risetime
1 ns
700 ps
(10%–90%, typical)
* Connected to oscilloscope with an input impedance of 1 MΩ ± 1%.
Electrical characteristics
Input resistance (system)
10 MΩ
± 1 %
Input capacitance (system)
9.5 pF
(typical)
Compensation range
10 pF to 25 pF
(typical)
Input coupling of the measuring instrument
1 MΩ AC / DC
Input impedance
Typical input impedance P2036 & P2056
10
100
1 k
10 k
100 k
1 M
10 M
100 M
10
100
1 k
10 k 100 k 1 M
10 M 100 M 1 G
Frequency (Hz)
|Z
| (Ω
)
CAUTION
To avoid overloading the circuit
under test, note that the input
impedance of the probe decreases
as the frequency of the applied
signal increases.
Mechanical characteristics
Weight (probe only)
48 g
Cable length
1.3 m
Probe tip diameter
2.5 mm