PicoSource PG900 software
28
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
4.4
Tunnel diode pulse head controls
TD Current
The
TD Current
controls how the instrument defines peak current of TD for selected
TD head.
·
Auto
- the instrument defines peak current automatically.
·
Manual
- the instrument controls peak current manually with the
Sensitivity
control. Use this mode for very fine tuning of TD.
Sensitivity
This controls peak current of TD manually when the
Manual
is selected from the
TD
Current.
Deskew
Set the deskew between
TD Head 1 Output
and
TD Head 2 Output
to compensate
for different cable delays. Set the
Deskew
between output in the range of 0 ps to
500 ps. Adjustable in 20 ps coarse increments or in 1 ps fine increments.
Summary of Contents for PicoSource PG900 Series
Page 2: ......