PicoSource PG900 Series User's Guide
27
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
On
You can switch on and off any of the
DRIVE
Outputs
with connected TD head.
TD Current
The
TD Current
controls how the instrument defines peak current of TD for selected
TD head.
·
Auto
- the instrument defines peak current automatically.
·
Manual
- the instrument controls peak current manually with the
Sensitivity
control. Use this mode for very fine tuning of TD.
Sensitivity
This controls peak current of TD manually when the
Manual
is selected from the
TD
Current.
Deskew
Set the deskew between
TD Head 1 Output
and
TD Head 2 Output
to compensate
for different cable delays. Set the
Deskew
between output in the range of 0 ps to 500
ns. Adjustable in a 20 ps coarse increment or in 1 ps fine increment.