pico ADC-200
installation guide
Version 2.2
Page 3
1.3 EMC
This instrument has been designed to meet the requirements of the EMC
Directive 89/336/EEC. Compliance was demonstrated by meeting the
test limits of the following standards :
Emissions
EN50081-1 (1992) Generic emission standard for residential, commercial
and light industry. Test methods and limits used were:
a) EN55022 Conducted, Class B.
b) EN55022 Radiated, ClassB.
Immunity
EN50082-1 (1992) Generic immunity standard for residential, commercial
and light industrial. Test methods and limits used were:
a) EN60801-2 (1993) Electrostatic Discharge.
b) IEC801-3 (1984) RF Field.
c) IEC801-4 (1988) Fast Transient.
pico ADC-200
installation guide
Page 4
Version 2.0
1.4 Specification
Parameter
ADC200-20
-50
-100
Max
sampling
rate
1 channel
20Msps
50 Msps
100Msps
2
channels
20Msps
50Msps
50Msps
Buffer size (samples)
2x8k
2x16k
2x32k
Resolution
8 bits
Analog Connections
2 channels
1MÙ input impedance
BNC connector
AC/DC coupling via switch (ADC 200-20)
AD/CD coupling via software (ADC200-50 and
ADC200-100)
Digital connections
1 channel: ext trigger/signal generator
>100kÙ input impedance
BNC connector
Voltage ranges
±20V, ±10V, ±5V, ±2V, ±1V, ±500mV, ±200mV,
±100mV, ±50mV
Accuracy
Voltage ±3%
Time ±100ppm
Trigger modes
Event: None/Rising/Falling
Source: chA/chB/Ext(digital)
Timing: pre/post/1% increments between
Output
D25 to PC printer port
Power supply
300mA @ 12V
500mA @ 12V
1.3mm DC connector (+ in centre)
Case
Grey ABS
140x190x45mm
not waterproof