•  The pressurization of the CUFF can temporarily cause loss of function of simultaneously used monitoring ME EQUIPMENT on 

the same limb.

•  Please use accessories and detachable parts specified/ authorised by the manufacturer. Otherwise, it may cause damage to the 

unit or danger to the user/patients.

•  We recommend checking its performance every two years or after it has been repaired. The device must be checked again 

in  case  of  damage  due  to  impact  (such  as  falls)  or  exposure  to  liquids  and/or  extreme  temperatures  (hot/cold)  or  extreme 

humidity variations.

•  For public place use, please contact your local distributor or the Italian toll-free number 800 900 080 to get support how to proceed.

•  Please dispose of accessories, detachable parts, and the ME EQUIPMENT according to the local guidelines.

•  In case of problems concerning settings, maintenance or use of the device, contact your local distributor or the Italian 

toll-free 

number 800 900 080.

 Warning! Do not service and/or repair the appliance while it is being used! Any other maintenance/

service activity/operation must be done solely by a PiC Service Centre. 

SAFETY INFORMATION

The symbols below can be found in the Instruction Manual as well as on labels and other components.

“PLEASE REFER TO THE INSTRUCTIONS 

FOR USE” symbol

“TYPE BF APPLIED PART” symbol

0068

Symbol for “COMPLIES WITH MDD 93/42/

EEC as amended by directive 2007/47/EC”

Storage and transport humidity symbol

“MANUFACTURER” symbol

Off symbol

SN

Symbol for “SERIAL NUMBER”

On symbol

 

   Latex free

Caution

Catalogue Number

The Bluetooth® combination mark, 

Bluetooth® version v. 4.0

0068

(2014/53/EU)

Symbol for “COMPLIES WITH RED 

2014/53/EU Directive”

Interference may occur in the vicinity of 

equipment marked with this symbol

Símbolo symbol for “ALTERNATE 

CURRENT”

Symbol for “DIRECT CURRENT”

IP22

Degree of protection against penetration of solids and liquids (dripping water when tilted 15°).

Appliance date of production: The second pair of digits in the lot number identifies the year, the third pair identifies the month. 

(e.g. LOT 1016100001: 16=2016; 10= October).

This section includes a list of error messages and frequently asked questions for problems you may encounter with your 

sphygmomanometer. If the products not operating as you think it should, check here before arranging for servicing.

PROBLEM

SYMPTOM

CHECK THIS

REMEDY

Error

message

No power

Display will not 

light up.

Low battery.

Recharge the batterys.

ON (I) - OFF (O) button 

switched OFF (O).

switch ON (I)  the ON (I) - OFF 

(O) button.

Low batteries

The display is dim or 

shows 

.

Low battery.

Recharge the battery.

E12 shows

Communication error

Check if the App is on or not, 

try data transmission again.

E01 shows

The cuff is not secure

Refasten the cuff and relax for a 

moment and then measure again.

E02 shows

The monitor detected 

motion, talking or the pulse is 

too poor while measuring.

Relax for a moment and then 

measure again.

E03 shows

The measurement process does 

not detect the pulse signal.

Loosen the clothing on the 

arm and then measure again.

E04 shows

The treatment of the measu-

rement failed.

Relax for a moment and then 

measure again.

EExx shows

A calibration error occurred.

Retake the measurement. If 

the problem persists, contact 

the retailer or our customer 

service department for further 

assistance. Refer to the war-

ranty for contact information 

and return instructions.

OUT

Out of range measurement.

Relax for a moment. Refasten 

the cuff  and then measure 

again. If the problem persists, 

contact your Physician.

SPECIFICATIONS

Power supply: 

   

 3.7V 1000mAh Built-in rechargeable Li-polymer battery (~120 measurements with full battery, 

battery rechargeable 300 times)

Display:  

Digital Backlight  LCD 25x86mm 

Measurement:

  

Oscillographic testing mode

Measurement range:

 

Rated cuff pressure: 0mmHg~300mmHg 

Measurement pressure:

 

SYS: 60-230mmHg, DIA: 40-130mmHg 

Pulse value:

   

40-199 beat/minute

Accuracy: 

 

 

Pressure: ± 3mmHg 

 

 

Pulse value: ±5% 

Normal working condition:

  Temperatures: from 5°C to 40°C 

 

 

Relative humidity≤85% 

 

 

Atmospheric pressure: 86kPa to 106kPa

Storage & transportation condition:

  Temperatures: -20°C +60°C 

 

 

Relative humidity 10%-93% 

Atmospheric Pressure:

 

50-106 kPa

Arm circumference:

  

About 22cm~42cm   

Net Weight: 

   

Approx. 295 g (including battery and cuff ) 

External dimensions:

  

Approx. 74.3x133.0x25.4mm (excluding cuff )

Mode of operation:

 

Continuous operation

Useful life:

 

 

10.000 Measurements 

Accessories:

    

AC/DC Adapter:

 

 100-240V~  50/60Hz  /  6V 

  1A  AC/DC  Adaptor  Model  KH0601000BW-PP-40  REF 

02009135000000(UK) or model KH0601000EW-PP-40 REF 02009136000000(EU) please refer to 

customer service

Degree of protection:

 

Type BF applied part (cuff )

Protection against ingress of water:

  IP22 degree of protection against penetration of solids and liquids (dripping water when tilted 15°)

Software version:

 

A01 

Device classification:  

Battery Powered Mode: Internally Powered ME Equipment

AC/DC Adaptor:

 

Class II Equipment

Contact Information

For  further  information  on  our  products,  please  visit  the  website  www.picsolution.com.  Here  you  can  find  customer  service, 

solve ordinary problems.

EMC Guidance

1)  mobileRAPID needs special precautions regarding EMC and needs to be installed and put into service according to the EMC 

information provided in the ACCOMPANYING DOCUMENTS

2) Portable and mobile RF communications equipment can affect mobileRAPID

3)  Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their 

base stations, walkie-talkies can affect this equipment and should be kept at least a distance d=3,3m away from the equi-

pment. The distance d is calculated by the manufacturer from the 800 MHz to 2.5 GHz column (Note: As indicated in Table 

6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell phone with a maximum output power of 2W yields d=3,3m at an 

IMMUNITY LEVEL of 3V/m).

4)  More information on EMC compliance of the mobileRAPID blood pressure monitor may be obtained from PIKDARE S.r.l using 

the contacts shown in this manual.

mobile

RAPID

M100% + Y100%

P 485 C

P 281 C

INSTRUCTIONS FOR USE

Automatic Blood pressure monitor 

PLEASE READ THESE INSTRUCTIONS CAREFULLY BEFORE USE AND KEEP 

THEM FOR FUTURE REFERENCE.

Helpful Tips

Here are a few helpful tips to help you obtain more accurate readings: 

• Do not measure your blood pressure immediately after consuming a large meal.

• To obtain more accurate readings, please wait one hour after meal before measuring.

• Do not smoke or drink tea, coffee or alcohol before measuring your blood pressure.

• Do not measure your blood pressure immediately after taking a bath: wait at least 20 minutes

• You should not be physically tired or exhausted while taking a measurement.

• Do not measure your blood pressure when you want to discharge urine

• It is important that you relax during measurement. 

• Do not take measurements if you are under stress or tense.

• Take your blood pressure at normal body temperature. If you are feeling cold or hot, wait a while before taking a measurement.

• Rest for 5 minutes before measuring.

• Wait at least 3 minutes between measurements. This allows your blood circulation to recover.

• Take the measurement in a silent room

• Do not move and do not talk when taking a measurement

• The cuff should maintain at the same level as the right atrium of the heart (Fig. A)

• Do not cross your legs and keep your feet on the ground.

• Keep your back against the backrest of the chair

•  For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately 

the same time, on the same arm, or as directed by a physician

Start the Measurement

If you want to store the measurement result on User1’s memory area, press the User1 button: it will automatically turn on, finish 

the whole measurement and and then save the result.

If you want to store the measurement result on User2’s memory area, press the User2 button: it will automatically turn on, finish 

the whole measurement and and then save the result.

1. 

When the monitor is off, press   or   button to turn on the monitor.

LCD  display                        

Zero setting        

Inflating

and measuring

Display and save

the results.

2. 

Press the User1 or User 2 button to power off, otherwise it will turn off within 1 minute.

If the   or 

 icon appears, please, read carefully the Helpful Tips section and repeat the measurement.

Notes: To stop the measurement at any time, press the   or   button under the display. The cuff will quickly deflate.

 CAUTION!

1.  When PiC mobileRAPID successfully pair-up with your SMARTPHONE, the measurement data will be auto-

matically transmitted to it via Bluetooth® wireless conntection.

WARNING!

Data transmitted and displayed on the terminal device cannot be used for direct diagnostic purposes . Please 

note that the blood pressure data transmitted to any smartphone (or tablet, etc.) is not intended to replace 

diagnosis or to give medical advices.

 

Recall the measurements

1.  When the Blood pressure monitor is off, press the “MEM” button to see the average of the last 3 recordings 

taken withn 30 minutes. If the recorded measurements are less than 3 as well as the last 3 recordings are 

taken in more than 30 minutes, you will see the latest reading.

2. The LCD will display the latest measuring result of the user ID which completes the last measurement.

The record number, measuring date and measuring time will be displayed alternatively. 

3. Press “Query” button to rotate the history records.

4.  When in the memory mode, press the User 1 button to recall the measurement history of User 1, or press the User 2 button to 

recall the measurement history.

5.  When no history stored for the specific user in the monitor, press “Query” button and the LCD will display as pictured to the right.

 CAUTION!

WARNING!

The most recent recording (1) is shown first. Every time a new recording is made, the first (1) record is assigned to it. All the others 

are moved by one figure (e.g. 2 becomes 3, and so on), whereas the last record (100) is deleted from the list.

Delete the measurements

(1) When under the query mode, press and hold “Query” button for 3 seconds to clear the memory.

(2) When the LCD display “dEL ALL”, press User button to confirm.

(3) The LCD will display “dEL doNE” and then shut off.

(4) Repeat points  from 1 to 3 pressing User 2 to delete the user 2 measurements

(5)  If you wish to stop clearing the memory, you may press the other user button to turn off the monitor, or wait until the monitor 

shuts off.

Maintenance

In order to get the best performance from your monitor, please follow the below instructions.

• Put in a dry place and avoid the sunshine.

• Avoid touching water, clean it with a dry cloth in case.

• Avoid the intense shaking and collision.

• Avoid the dusty and unstable temperature environment.

• Using the wet clothing to remove the dirt.

• Avoid washing the cuff.

Please use the soft cloth to clean the whole unit. Don’t use any abrasive or volatile cleaners.

Calibration and service

The accuracy of this monitor has been carefully tested and is has been designed for a long life. Indeed the monitor can maintain 

the safety and performance characteristics for a minimum of 10,000 measurements or five years of normal use.

It is generally recommended to check the unit every two years to ensure correct functioning and accuracy. Please consult the 

Contact information in this user manual to get in touch with your local authorized Service Center.

For battery or cuff replacement contact exclusively the authorised Pic Technical Customer Service.

WARNINGS

• This unit is not suitable for continuous monitoring during medical emergencies or operations.

• This device cannot be used with HF surgical equipment at the same time.

• This device is not intended for patient transports outside a healthcare facility.

• This device is intended for adult use only. Keep out of reach of children.

• People with reduced cognitive abilities shall use the device under supervision.

• This device is intended for no-invasive measuring and monitoring of arterial blood pressure.

•  It is not intended for use on extremities other than the arm or for functions other than obtaining a blood pressure measurement.

•  Do not confuse self-monitoring with self-diagnosis. By using this device you can monitor your blood pressure, but it does not 

replace your doctor. Begin or end medical treatment based solely on medical advice. If you are taking medication, consult your 

physician to determine the most appropriate time to measure your blood pressure. Never change a prescribed medication 

without consulting your Physician.

•  If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the procedure of blood pressure measurement, a 

signal of will be displayed. Under this condition, the Electronic Sphygmomanometer can keep function, but the results may not 

be accurate, it’s suggested that you consult with your physician for accurate assessment

•  When the device is used for measurements on patients with common arrhythmias, such as atrial or ventricular premature beats 

or atrial fibrillation, a deviation may occur. Please consult your doctor about the result. If the arm cuff pressure exceeds 300 

mmHg it deflate automatically. If the arm cuff does not deflate when pressure exceeds 300 mmHg, take it off your arm and press 

the   or   button to stop inflation.

•  The device is a non-AP/APG device and not suitable for use in the presence of anaesthetic mixtures that are inflammable with 

air, oxygen or nitrous oxide.

•  The operator shall not touch DC output jack of AC/DC adapter and the patient simultaneously.

•  To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated interference signal or electrical 

fast transient/burst signal.

•  The user must check that the equipment functions safely and see that it is in proper working condition before being used.

•  This device is contraindicated for any female who may be suspected of, or is pregnant. Besides provided inaccurate readings, the 

effects of this device on the fetus are unknown.

•  Manufacturer will make available on request circuit diagrams, component parts list etc. Only to qualified person.

•  Too frequent measurements can cause injury to the PATIENT due to blood flow interference. Please  check (by observation of the limb 

concerned) that operation of mobileRAPID does not result in prolonged impairment of the circulation of the blood of the PATIENT.

•  This unit is not suitable for continuous monitoring during medical emergencies or operations. Otherwise, the patient’s arm and 

fingers will become anaesthetic, swollen and even purple due to a lack of blood.

•  Please use the device under the environment which was provided in the instruction manual. If this is not complied with, correct 

operation and life of the device may be compromised.

•  During using, the patient will contact with the cuff. The materials of the cuff have been tested and found to comply with require-

ments of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction..

• The application of the CUFF over a wound, as this can cause further injury

•  The application of the CUFF and its pressurization on any limb where intravascular access or therapy, or an arterio-venous (A-V) 

shunt, is present because of temporary interference to blood flow and could result in injury to the PATIENT

•  The application of the CUFF and its pressurization on the arm on the side of a mastectomy can cause injury or give not accurate 

measurement

SYMBOL

DESCRIPTION

EXPLANATION

Systolic pressure

Maximum/systolic pressure result

Diastolic pressure

Minimum/diastolic pressure result

Pulse

Pulse/minute

Deflating

The arm cuff is deflating

mmHg

Blood pressure unit of measurement mmHg

Low battery

The battery is low and needs recharging

Irregular heartbeat

Irregular heartbeat

Grade

Blood pressure level indicator

Date

“M” shows the month, “D” shows the day

Average 

The average of last 3 measurements taken within 30 minutes.

Data transmitting

Data transmission succeeds (Bluetooth® functional icon)

Memory

If     shows, the displayed measurement values is from the memory

Heartbeat

Heartbeat Detection during the measurement

Motion detected while 

measuring

Measure could be inaccurate

The steps that the patient/user/operator can do safely are only those specified in the following instructions of use, maintenance 

and warnings. 

Recharge the device battery

1. The battery of the mobileRAPID is a built-in rechargeable Li-polymer battery type, the battery is 3.7V 1000 mAh.

2.  Please use only the AC/DC adaptor Model: KH0601000BW-PP-40 REF 02009135000000(UK) or model KH0601000EW-PP-40 REF 

02009136000000(EU) to charge the battery, just like the following picture:

AC Adaptor

Charging the power under following circumstances:

• 

 Is displayed , 

• The LCD display dims, 

• When powering on the monitor, the display does not light up. 

 CAUTION!

WARNING!

  Use  only  the  PiC’s  authorized  AC/DC  Adaptor  (KH0601000BW-PP-40  REF  02009135000000(UK)  or  model 

KH0601000EW-PP-40REF  02009136000000(EU)  to  charge  the  power. You  cannot  use  the  blood  pressure  monitor  during  the 

process of charging.

WARNING!

•  Pic mobileRAPID has a built-in rechargeable Li-polymer battery, and cannot to be replaced  or disassembled by the customer 

and unauthorized maintenance personnel. 

•  The  battery  must  be  disassembled  and  replaced  only  by  PiC  authorized  maintenance  personnel.  Do  not  try  to  change  the 

battery by yourself: improper substitution or the use of  wrong spare part can cause risk of explosion. 

•  Under the normal using, it can charge power about 300 times, if the battery cannot charge the power normally or the blood 

pressure monitor cannot use normally, please connect with the authorized maintenance personnel. If measured three times per 

day, and the battery is fully charged, it can be used for about 40 days.

•  In order to extend the battery charge, please switch off the device after usage.

•  Storage and use the blood pressure monitor at the cool, dry and ventilated environment. 

•  

Caution!

 Avoid to approach to the fire and the heat source, or it will cause the battery explode. The battery shall be disposed 

according to the guidline shared at the end of this manual (2006/66/EC)

•  During the process of charging, the blood pressure monitor displays :  

•  When the charging is finished, please pull the plug.

•  When charging, shall not touch charging connector and the patient simultaneously.

•  When charging with adaptor, the socket-outlet shall be installed near the equipment and shall be easily accessible.

•  Lithium batteries require charging prior to use. Always refer to the instructions for correct charging procedures.

•  Do not keep charging the lithium battery if fully charged.

Setting Date and Time

To ensure the stored measurement result has correct time record, please set time and unit before device is used. 

Before use, switch the ON/OFF button to the “I” side to turn on the monitor. Note: If the button is on the “ ” side, there is no 

reaction when you press any button.

(1) When the monitor is off press and hold User 1 button for 3s to enter Time Setting Mode.

(2)  The year will flash on the monitor, press “Query” button to change the number Each press will increase the numeral by one 

in a cycling manner. 

(3)  Press “User 1” button again to confirm [YEAR]. Then the number representing the [MONTH] blinks.

Press “Query” button to change the number. Each press will increase the numeral by one in a cycling manner.

Representing the [DAY] blinks. Press “Query” button to change the number. Each press will increase the numeral by one in a 

cycling manner.

(4)  Press “User 1” button again to confirm [DAY] Then the number representing [24 or 12h mode] blinks. Press “Query” button to 

change the number. Each press will change the numeral by 12 to 24h in a cycling manner.

(5)  Press “User 1” button again to confirm [24 or 12h mode] Then the number representing [HOURS] blinks. Press “Query” button 

to change the number. Each press will increase the numeral by one in a cycling manner.

(6)  Press “User 1” button again to confirm [HOURS] Then the number representing [MINUTE] blinks. Press “Query” button to change 

the number Each press will increase the numeral by one in a cycling manner.

(7)  Press “User 1” button again to confirm [MINUTE] Then the LCD will display “doNE” 

 YEAR 

 DAY & MONTH 

 24 or 12h 

MODE   HOUR & MINUTE then the monitor will shut OFF.

Note: If during the setting mode no button is pressed for more of about 60s the display  turned OFF automatically.  

Pairing up the Blood Pressure Monitor with Your Device

(1) Turn on Bluetooth® conntection and the app Pic Health Station. Make sure both are on when pair-up is proceeding.

(2)  When the monitor is off, press and hold the User 2 button to start pair-up. The symbol   and the symbol   will be shown 

on the LCD alternatively, indicating pair-up is proceeding. If succeed, symbol   will be shown on the LCD. If fail, symbol   

will be shown on the LCD.

(3) The monitor will shut off automatically after pair-up process is complete.

(4)  During the pairing, only the smartphone to be connected with the device shall run PIC APP. If you meet problem in pairing 

please check that no other smartphone in the family are running the APP, or reboot your smartphone.

Bluetooth® Module No. : nRF51822

Frequency Range: 2.402 - 2.480 GHz

Supply Voltage: 3.3 V

Output Power Range: 4 dBm

Applying the Cuff

1.  Remove all watches, jewellery, etc. prior to attaching the arm monitor. Clothing sleeves should be rolled up and the cuff should 

be wrapped on bare skin for correct measurements.

2. Apply cuff preferably to left arm with palm facing up as Fig. A.

3. Make sure the edge of the cuff is about 2-3 cm from the elbow.

4.  In order to ensure accurate measurements, fasten the velcro strap securely around your arm so there is no extra space between 

the cuff and your skin. If the cuff is too loose, the measurement will not be accurate.

Fig. A

THESE INSTRUCTIONS ARE IMPORTANT. PLEASE KEEP THEM FOR FUTURE REFERENCE. 

Dear customer, 

Thank you for choosing Pic Solutions mobileRAPID, an automatic blood pressure monitor designed and manufactured according 

the most up-to-date technologies. Before using the device, we highly recommend that you read the short list of warnings con-

tained in this manual, to make sure you have correctly understood how to use it. 

INTRODUCTION 

The  regular  use  of  a  Blood  Pressure  Monitor  allows  you  to  monitor  blood  pressure  continuously,  but  it  should  not  take  the 

place of regular check-ups by your doctor. We recommend that you visit your doctor regularly for general check-ups and for 

more detailed information on blood pressure. Blood pressure measurements determined with this unit are equivalent to those 

obtained by a trained observer using cuff/ stethoscope auscultation method, within the limits prescribed by the international 

ESH Validation protocol. 

WARRANTY 

The product is guaranteed for 5 years against all conformity defects under the normal utilisation conditions provided in the 

user instructions. Therefore, the warranty shall not apply in case of damages caused by improper use, wear, or accidental events. 

Batteries, cuff and wearing parts are not included in the warranty. In case of malfunction, contact exclusively the authorised Pic 

Technical Customer Service. 

RAPID-TECH Comfort Experience

TM 

This unit uses the oscillometric method to detect your blood pressure. RAPID-TECH Comfort Experience

TM

 is the latest generation 

technology chosen by Pic for the blood pressure monitors of the line RAPID. It allows fast and, therefore, less annoying monito-

ring because the measurement is performed already during the armband inflation phase. Pic mobileRAPID has been clinically 

validated in accordance with the ESH (European Society of Hypertension) protocol, as a guarantee of its accuracy and precision. 

DC power socket

Systolic pressure

Cuff (Type BF applied part)

Query/Up button

User 1/Set button

Diastolic pressure

ON/OFF button

User  2/Activation button

Pulse rate

List of included items: Monitor with the M-L size cuff, mains adaptor, user manuals and carring bag. 

Features: 

• 24.5x86.5 mm Blue LCD display with blue backlight 

• Two memory areas 

• 100 recorded measurement per area 

• Inflation-type measurement

• Wireless connectivity (Bluetooth® version v. 4.0)

• App available for iOS and Android

ABOUT BLOOD PRESSURE

What is blood pressure?

Blood pressure is the pressure exerted on the artery tube while blood flows through the arteries. The pressure measured when 

the heart contracts and sends blood out of the heart is systolic (highest) blood pressure. The pressure measured when the heart 

dilates with blood flowing back into the heart is called diastolic (lowest) blood pressure. 

Why measure your blood pressure? 

Among  the  various  health  problems  afflicting  modern  people,  problems  associated  with  high  blood  pressure  are  by  far  the 

most common. High blood pressure’s dangerously strong correlation with cardiovascular diseases and high morbidity has made 

measuring blood pressures a necessity of identifying those at risk. 

Blood Pressure Fluctuation

Blood pressure fluctuates all the time! You should not be over worried if you encountered two or three measurements at high levels. 

Blood pressure changes over the month and even throughout the day. It is also influenced by the season, temperature and activities. 

Classification of high blood pressure

The out-of-office blood pressure classification published by ESH/ESC in 2013 is as follows: 

Note:

  It  is  not  intended  to  provide  a  basis  of  any  type  of  rush 

toward  emergency  conditions/diagnosis  based  on  the  scheme: 

the scheme is meant only to discriminate between the different 

levels of blood pressure.

Hypertension

Normal Blood Pressure

Systolic blood pressure (mmHg)

Diast

olic blood pr

essur

e (mmHg)

 CAUTION 

Only a physician can tell your normal BP range. Please contact a physician if the average of your measuring results falls out of the 

range. Kindly note that only a physician could tell whether your blood pressure value has reached a dangerous point.

Irregular Heartbeat Detector

This blood pressure monitor is equipped with an algorithm for Irregular Heartbeat (IHB) Detection. During each measurement, 

this algorithm records the heartbeat intervals and works out the standard deviation. If the calculated value is larger than or equal 

to 15, this algorithm will light up the IHB symbol on the screen when displaying the measuring result.

 CAUTION  

The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heartbeat was detected d uring mea-

surement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. 

We wish to underline that the device does not replace a visit to the cardiologist, but it is useful to detect any irregularities in advance.

 Why the blood pressure I get from the hospital is different from home? 

The blood pressure is different even during 24 hour because of the weather, emotion, exercise etc, specially the “white coat” in 

hospital which makes the results are higher than the ones at home. 

If the result is the same if measuring on the right arm?

Measurements can be taken on either arm; results can however differ sometimes so it is always advisable to use the same arm.

LCD display explanation

At starting check the completeness of the LCD segments

High

Normal

Blood pressure

grade indicator

Summary of Contents for Mobile Rapid

Page 1: ...stato di gravidanza Oltre a determinare letture imprecise non si conoscono gli effetti dell apparecchio sul feto Su richiesta il produttore fornir gli schemi dei circuiti l elenco dei componenti ecc...

Page 2: ...per radiotelefoni cellulari cordless e radiomobili terrestri apparecchi per radioamatori trasmettitori radio in AM e FM e trasmettitori TV non possono essere previste teoreticamente con precisione Per...

Page 3: ...trong electromagnetic field radiated interference signal or electrical fast transient burst signal The user must check that the equipment functions safely and see that it is in proper working conditio...

Page 4: ...ation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the mobileRAPID b Over the frequency range 150 kHz to 80 MHz field strengths should be...

Page 5: ...h mas des circuits la liste des composants etc uniquement aux personnes qualifi es Des mesures trop fr quentes peuvent provoquer des l sions au PATIENT du fait de l interf rence avec le flux sanguin V...

Page 6: ...os amateurs la radiodiffusion AM et FM et la diffusion t l vis e ne peuvent pas tre th oriquement pr dites avec pr cision Afin d valuer l environnement lectromagn tique d aux metteurs RF fixes une tud...

Page 7: ...hten Messergebnissen kommen Bitten Sie Ihren Arzt um Hilfe bei der Auslegung der Messwerte Steigt der Druck der Manschette ber 300 mmHg entl ftet sich das System automatisch Sollte die Manschette sich...

Page 8: ...etische Standortaufnahme fest gelegt a niedriger sein als die Konformit ts stufe f r jeden Frequenzbereich b St rungen k nnen in der N he von mit dem folgenden Symbol gekennzeichneten Ger ten auftrete...

Page 9: ...a la circulaci n sangu nea del PACIENTE Este aparato no es apto para la monitorizaci n continua en caso de emergencias o intervenciones m dicas En tal caso el brazo y los dedos del paciente podr an en...

Page 10: ...res emisoras de aficionados emisoras de radio AM y FM y emisoras de televisi n no se pueden predecir te ricamente con exactitud Para evaluar el entorno electromagn tico provocado por transmisores de r...

Page 11: ...o do aparelho contraindicada em mulheres gr vidas ou que suspeitem estar gr vidas Al m de determinar leituras imprecisas os efeitos do aparelho sobre o feto n o s o conhecidos O fabricante pode forne...

Page 12: ...o podem ser previstas teoricamente com exatid o Para a avalia o do ambiente eletromagn tico devido a transmissores de RF fixos deve ser considerado um levantamento eletromagn tico do local Se a inten...

Reviews: