Discharging a P-736.ZRx connected to the controller
In closed-loop operation:
1.
Switch off the servo mode on the controller.
2.
Set the piezo voltage to 0 V on the controller.
In open-loop operation:
Set the piezo voltage to 0 V on the controller.
Discharging a P-736.ZRx not connected to the controller
Connect the P-736.ZRx to the switched-off PI controller for 10 seconds.
Maintenance
NOTICE
Damage due to improper maintenance!
The P-736.ZRx is maintenance-free and precisely aligned.
Loosen any screws only when instructed in this manual.
Do
not
open the P-736.ZRx.
Cleaning the P-736.ZRx
NOTICE
Damage from ultrasonic cleaning!
Ultrasonic cleaning can damage the
P-736.ZRx.
Do
not
do any ultrasonic cleaning.
Requirements
You have discharged (p. 17) the piezo actuators of the P-736.ZRx.
You have disconnected the P-736.ZRx from the controller.
Cleaning the P-736.ZRx
Clean the surface of the P-736.ZRx with a cloth that is dampened with a mild cleanser or
disinfectant (e.g., alcohol or isopropyl alcohol).