Connecting the P-736.ZRx to the Controller
Requirements
You have read and understood the safety precautions (p. 6).
The controller is switched off, i.e.,
not
connected to the power source.
You have connected the P-736.ZRx to the protective earth conductor (p. 11).
Connecting the P-736.ZRx to the controller
Connect the D-sub 9 connector of the P-736.ZRx to the corresponding D-sub 9 socket of
the controller.
Startup and Operation
Starting and Operating the P-736.ZRx
INFORMATION
Systems are calibrated at the factory to achieve optimum performance. Replacing the system
components will cause a loss in performance.
Note the assignment of the stage axes to the controller channels, which is specified by the
calibration label of the piezo servo controller.
If the piezo servo controller or the stage has to be replaced, recalibrate the axis
displacement (refer to the controller manual) or contact our customer service department
(p. 19).
Requirements
You have read and understood the safety precautions (p. 6).
You have connected the P-736.ZRx to the protective earth conductor (p. 11) and to the
controller (p. 17).
Starting and operating the P-736.ZRx
Follow the instructions in the user manual for the piezo controller used for startup and
operation of the P-736.ZRx.
Discharging the P-736.ZRx
The P-736.ZRx must be discharged before demounting. Demounting is necessary e.g., before
cleaning or transporting the P-736.ZRx as well as for modifications.
Requirements
You have read and understood the safety precautions (p. 6).