5 Installation
M-232 Linear Actuator
MP105E
Version: 1.2
23
2. Unscrew the grub screw in the middle of the front of the micropositioning
stage by around three rotations using an Allen wrench or screwdriver.
Figure 7: Unscrew the grub screw slightly
3. Align the actuator so that the following conditions are met:
−
The locating face of the actuator is level with the top edge of the
micropositioning stage.
−
The mounting shaft is located precisely across from the horizontal hole on
the front of the micropositioning stage.
Figure 8: Aligning the actuator