4
www.phywe.com, © All rights reserved
12933-01 / 4521
Take the temperatu
re and atmospheric pressure into
consideration.
hPa:
mmHg:
1026,5
770
1013,3
760
999,9
750
986,6
740
973,3
730
16°C
10.07
9.94
9.81
9.68
9.55
17°C
9.86
9.74
9.61
9.48
9.35
18°C
9.67
9.54
9.41
9.29
9.16
19°C
9.47
9.35
9.23
9.11
8.98
20°C
9.29
9.17
9.05
8.93
8.81
21°C
9.11
9.00
8.88
8.76
8.64
22°C
8.94
8.83
8.71
8.59
8.48
23°C
8.78
8.66
8.55
8.44
8.32
24°C
8.62
8.51
8.40
8.28
8.17
25°C
8.47
8.36
8.25
8.14
8.03
26°C
8.32
8.21
8.10
7.99
7.89
27°C
8.17
8.07
7.96
7.86
7.75
28°C
8.04
7.93
7.83
7.72
7.62
29°C
7.90
7.80
7.69
7.59
7.49
30°C
7.77
7.67
7.57
7.47
7.36
31°C
7.64
7.54
7.44
7.34
7.24
32°C
7.51
7.42
7.32
7.22
7.12
33°C
7.39
7.29
7.20
7.10
7.01
34°C
7.27
7.17
7.08
6.98
6.89
35°C
7.15
7.05
6.96
6.87
6.78
hPa:
mmHg:
959,9
720
946,6
710
933,3
700
919,9
690
906,6
680
16°C
9.42
9.29
9.15
9.02
8.89
17°C
9.22
9.10
8.97
8.84
8.71
18°C
9.04
8.91
8.79
8.66
8.54
19°C
8.86
8.74
8.61
8.49
8.37
20°C
8.69
8.57
8.45
8.33
8.20
21°C
8.52
8.40
8.28
8.17
8.05
22°C
8.36
8.25
8.13
8.01
7.90
23°C
8.21
8.09
7.98
7.87
7.75
24°C
8.06
7.95
7.84
7.72
7.61
25°C
7.92
7.81
7.70
7.59
7.48
26°C
7.78
7.67
7.56
7.45
7.35
27°C
7.64
7.54
7.43
7.33
7.22
28°C
7.51
7.41
7.30
7.20
7.10
29°C
7.39
7.28
7.18
7.08
6.98
30°C
7.26
7.16
7.06
6.96
6.86
31°C
7.14
7.04
6.94
6.85
6.75
32°C
7.03
6.93
6.83
6.73
6.63
33°C
6.91
6.81
6.72
6.62
6.53
34°C
6.80
6.70
6.61
6.51
6.42
35°C
6.68
6.59
6.50
6.40
6.31
5.5 Care
Membrane
Ensure not to damage the membrane and follow the storage
instructions in order to increase the service life of the elec-
trode. Creases or holes in the membrane are a sign that it
does not work properly any more.
If the membrane is soiled or if it shows signs of damage (this
might be indicated by incorrect measurement values), use a
spare cap with a membrane.
Cathode and anode
They must be polished with polishing cloths from time to time.
To do so, remove the cap so that the anode and cathode can
be accessed. Polish them with slight pressure and a clean
cleaning cloth in order to remove any deposits from the sur-
face. Rinse them with distilled water and then with alcohol,
shake the liquid off, and screw the cap back on loosely.
6 TECHNICAL DATA
Operating temperature range: 5 - 40°C
Rel. humidity < 80%
Oxygen in air:
Measuring range
Resolution
Accuracy
Oxygen in liquid:
Measuring range
Resolution
Accuracy (10~35°C)
0…100 %
0.1 %
±2%
0…20 mg/l
0.01 mg/l
±0.5mg/L
Max. date rate
Battery capacity
Max. wireless range (open field)
10 Hz
250 mAh
30 m
Dimensions (length x width x height)
85 x 40 x 23 mm
Weight
40 g
7 SCOPE OF DELIVERY
The extent of delivery is as follows
•
Cobra SMARTsense Oxygen
12933-01
•
USB connecting cable type C
07935-00
•
Dissolved and Air Oxygen Electrode
•
Operating instructions
8 ACCESSORIES
The following accessories are available:
•
Cobra SMARTlink
12999-99
•
USB-charger
07934-99
•
USB connecting cable type C
07935-00
•
USB-Bluetooth-Adapter
07936-00
•
Calibration solution Zero Oxygen, 500 ml
47072-50
•
Software measureLAB
14580-61
•
Free measureApp available from supplier portals
iOS
Android
Windows