background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

12933-01 / 4521

 

 

1.

  Unscrew the lower end

  

 

 
 
 
 
 
 
 

2.  Fill the cap with the liquid  

 

 
 
 
 
 
 
 
 

3.  Screw the filled cap back onto the electrode. Use a 

cloth to wipe off any liquid that leaks out. 

Preparation 

Connect  the  oxygen  probe  (included  in  the  scope  of supply) 
to  the  connector  at  the  front  of  the  sensor.  Ensure  that  you 
can  hear  that  the  outer  metal  ring  of  the  connector  locks  in 
place. 

ATTENTION:

  Never  turn  the  plug  after  it  has  been  plugged 

in. This can lead to the destruction of the sensor. 

 

Calibration 

It  is  not  necessary  to  recalibrate  the sensor  for  every meas-
urement. Especially if only the change in oxygen content is to 
be demonstrated, an already existing calibration (the last cal-
ibration  remains  automatically  stored  in  the  sensor)  or  the 
factory  settings  are  usually  sufficient.  However,  if  absolute 
values are required for your measurement, e.g. if you are de-
termining  the  oxygen  content  of  bodies  of  water,  or  if  you 
want to optimise the accuracy of your measurement, we rec-
ommend calibrating the sensor. 

See section 5.4

 

5.3 Recording of measurement data 

 

Measurement: 
Two  types  of  measurements  can  be  performed  with  the  Co-
bra SMARTsense Oxygen: Measurement of the oxygen con-
tent,  e.g.  in  the  air  (%),  and  a  measurement  of  the  oxygen 
concentration in an aqueous solution (mg/l).  
a) 

Measurement of oxygen in the air 

  Select the "Air" mode in the software. 
  It takes approximately 1 minute for the value to settle. 

b) 

Measurement of dissolved oxygen 
Select the "Fluid" mode in the software. 

 

Plunge  the  oxygen  probe  into  the  solution  so  that 
the  temperature  sensor  is  also  covered  (approxi-
mately.  5-6  cm).  Do  not  immerse  it  completely  in 
the liquid. The handle is not waterproof. 

 
 
 
 
 
 

 

Record  the  oxygen  concentration  with  the  aid  of 
the recording function of the software. 

 

Ensure  that  the  water  sample  that  is  to  be  ana-
lysed  is  slightly  stirred  in  a  continuous  manner. 

This  is important since the  sensor  consumes oxy-
gen  during  the  measurement.  Without  stirring,  it 
would  seem  as  if  the  oxygen  concentration  de-
creases. 

 

Measuring principle: 
This sensor type is an amperometric Clark sensor with a plat-
inum cathode and a silver anode.  
A  fixed  voltage  is  applied  at  the  platinum  electrode.  When 
oxygen  diffuses  through  the  membrane  and  to  the  cathode, 
the latter will be reduced:  
 
1/2 O

2

 + H

2

O + 2

e

–  

→ 

2 OH

 

 
The reference electrode (anode) is subject to oxidation:  
 
Ag + Cl

 

→ 

AgCl + e

– 

 
Correspondingly,  a  current  flows  and  this  current  is  propor-
tional  to  the  oxygen  diffusion  and,  thereby,  to  the  oxygen 
concentration.  The current  is converted  into a  voltage signal 
that is then received by the sensor. 

 
5.4 Calibration 

Ensure to perform the calibration at a temperature that is sim-
ilar  to  the  temperature  of  the  sample.  Like  in  the  case  of  a 
measurement,  wait  for  approximately  1-2  minutes  after  the 
connection  of  the  sensor  until  the  measurement  value  has 
settled. Then, perform the calibration. 
 

Oxygen in air 

A two-point calibration is performed.

 

Select the "Air" mode. 
 
Use a commercial zero-oxygen calibration solution. 

 

 
First calibration point 
Select  "Configuration/Calibration"  from  the  software  menu 
(APP). Immerse the measuring probe into the calibration liq-
uid and wait approx. 2 minutes until the measurement values 
have settled. Enter the 0% value as set point 1 and send it to 
the sensor. 
 
Second calibration point 
Hold  the  sensor  in  the  ambient  (outdoor)  air  since,  in  this 
case,  the  oxygen  content  will  be  constantly  at  20.9%.  Wait 
approx. 2 minutes until the measurement values have settled. 
Enter the 20.9% value as set point 1 and send it to the sen-
sor. 

 

Dissolved oxygen 

A two-point calibration is performed. 
 
Use a comercial zero-oxygen calibration solution. 

 

 
First calibration point 
Select  "Configuration/Calibration"  from  the  software  menu 
(APP). Immerse the measuring probe into the calibration liq-
uid and wait approx. 2 minutes until the measurement values 
have settled. Enter the 0 mg/l value as set point 1 and send it 
to the sensor. 
 
Second calibration point 
Hold the oxygen probe into a receptacle fill with distilled wa-
ter.  The  measurement  value  should  have  settled  after  ap-
prox.  two  minutes  and  the  calibration  can be performed.  Set 
point 2 can be found in the table below.  

Temperature sensor 

Reviews: