1.0 GENERAL 

DESCRIPTION

1.1

The TCAT-2 is a general purpose autotuning controller for a variety of laboratory applications. 

The TCAT-2AC provides two time-proportioning controlled AC outlets for the control of resistive
loads such as heat lamps or warming blankets. It can also be adapted to many different types of load
(maximum 500 Watts.)

The TCAT-2LV provides a low voltage DC output for electrically sensitive applications where AC fre-
quencies will interfere with monitoring devices. It can be supplied with a variety of different DC
power supplies depending on the load requirements. The TCAT-2LV uses a dual set point control cir-
cuit to prevent a heating element from overshooting the control point. This is recommended in applica-
tions such as animal warming tables or blankets where thermal lag in the control system is an issue and
might lead to burns.

The TCAT-2DF combines features of both the  instruments described above. 

A type T thermocouple probe is used as the temperature feedback sensor and any of Physitemp’s
extensive line of probes can be used for this application. 

All the TCAT-2 instruments operate on a 110V - 230V AC supply.

2.0

UNPACKING AND CHECKING THE PACKAGE CONTENTS

The following items should be found in the packing box:

TCAT-2 controller
AC Line cord
AC Jumper cable
Probes and other items as ordered.

2

QX P:\CW shared\TCAT\EB TCAT-2DF PAK 400 Rev 2     3/5/13

Summary of Contents for TCAT-2DF

Page 1: ...ENTS 3 0 SETTING UP THE TCAT 2DF 3 4 0 OPERATING INSTRUCTIONS 4 5 0 SPECIFICATIONS 5 6 0 MAINTENANCE WARRANTY AND 6 SERVICE Installation instructions for HP 1 and HP 4 Warming Plates Installation inst...

Page 2: ...nts The TCAT 2LV uses a dual set point control cir cuit to prevent a heating element from overshooting the control point This is recommended in applica tions such as animal warming tables or blankets...

Page 3: ...ations Department 1 800 452 8510 for more information on sensor placement 3 3 Connect the AC power cord to the AC input module and to any suitable AC outlet 3 4 The auxiliary AC output socket is avail...

Page 4: ...o change the flashing digit Press SHIFT again to select the next digit c When all digits are changed press SET to enter new control temperature SPECIFICATIONS Operating Range 200 C to 400 C Resolution...

Page 5: ...ectable Fused input module 6 3Amp 5 x 20mm fuse Output Via LVDC power supply or 2 controlled auxiliary AC sockets 500W AC max Size 7 3 8 wide x 10 deep x 4 3 8 high Weight 2 8 lbs Other features Speci...

Page 6: ...ument contact us at Tel 1 800 452 8510 or 1 973 779 5577 Fax 1 973 779 5954 E mail info physitemp com In the event that any part of this system is to be returned for repair or recalibration please pac...

Page 7: ...e control temperatures are required please call our service department for instructions on how to change the set point 4 A second sensor Type T thermocouple described in the Operating Manual for TCAT...

Page 8: ...maximum for WB 2 Efficiency of output power 98 OPERATING PROCEDURE 1 Connect the WB 1 or WB 2 extension lead to the rear of the TCAT 2LV or TCAT 2DF con troller supplied with it Do not connect to any...

Page 9: ...C 154 Huron Avenue Clifton NJ 07013 USA Tel 1 800 452 8510 Tel 1 973 779 5577 Fax 1 973 779 5954 e mail info physitemp com www physitemp com OPERATING MANUAL TCAT 2DF CONTROLLER QX P CWS TCAT EB TCAT...