Connect AC line cord from the controller to an AC outlet and turn ON. 
 
A buzzing noise from water pump indicates that pump is running. Remove tank cap and 
check that water is flowing. 
 
The power indicator light on the controller and tank unit will light. IF EITHER  LAMP 
FAILS TO LIGHT, SWITCH OFF and check fuses. 
 

3.0 OPERATING 

INSTRUCTIONS 

 

3.1 

ADJUSTMENT OF BASE TEMPERATURE OF CONTROLLER 

 
All parts of the system must be connected and turned on. 
 
Set all switches under SET DIFFERENTIAL TEMPERATURE to center  '0' position. 
 
Depress RUN/SET switch to SET position and hold down while adjusting SET BASE 
TEMP with screwdriver. Adjustment may be made to any base temperature between 20

 and 

30

 Centigrade to within 0.1

 C. 

 
Release the RUN/SET switch and allow the stage to stabilize at the base set temperature. 
This should take one to two minutes. Further slight adjustments to temperature may be made 
to provide exact setting. 
 
All further temperature settings can be made using the digital control settings (switches on 
the front panel). These settings allow the temperature of the stage to be incremented or 
decremented in 5

, 1

 degree or 1/10

 steps from the base temperature. 

 

3.2

 

MEASURING ABSOLUTE TEMPERATURE 

 

The digital display of the controller may be used as a readout for an external sensor (Type T, 
copper-constantan thermocouple with miniature t/c plug.) 
 
Connect a thermocouple sensor into socket marked SENSOR A - EXTERNAL. 
 
Switch SENSOR INPUT to A. Read absolute temperature on controller display. 
 
During measurement with external sensor, all power to the Thermal Probe is disconnected.  
Switch SENSOR INPUT back to B and allow the stage temperature to stabilize again. 
 

3.3 WARNING 

LIGHT 

 
As a protective feature, both audible and visual warnings of operating failure are provided. 
 
SWITCH OFF IMMEDIATELY IF WARNING LIGHT COMES ON OR IF BUZZER 
ALARM IS HEARD. 
 
In the event of failure, consult manufacturer for further information:  
                Physitemp Instruments Inc.  973-779-5577 or e-mail [email protected]

 

4.0 SPECIFICATIONS 

 

Summary of Contents for NTE-2A

Page 1: ...t Supplied 2 2 Water Pump and Tank Unit 2 3 Attaching AC Line Cords 2 4 Attaching Water Tubes 2 5 Connecting the Thermal Probe 2 6 Connection to AC and Switch On 3 0 OPERATING INSTRUCTIONS 3 1 Adjusting Base Temperature of Controller 3 2 Measuring Absolute Temperature 3 3 Warning Light 4 0 SPECIFICATIONS 5 0 WARRANTY AND SERVICE W Shared docs Operating manuals EB NTE 2A Rev 1 01 25 07 ECO 10292 ...

Page 2: ...nd below room temperature The temperature of the tip of probe can be adjusted using the SET DIFFERENTIAL TEMP switch of the controller which maintains the temperature at the set point The digital readout on the controller continuously indicates the actual tip temperature of the probe For stable operation the thermoelectric module requires a trickle of cooling water This is supplied from the Pump a...

Page 3: ...lowing when the controller is turned on Check water level periodically during operation to ensure that water level is within three inches of the lip of the container To prevent spills all water connections are automatically self sealing when disconnected The pump in the PTU 110A Pump and Tank Unit is self priming When first started it will be noisy for five or ten seconds This is normal After the ...

Page 4: ...ut and reinserted with the proper orientation Please note The controller can be set to 120V or 240V However the pump and tank unit does not have voltage setting adjustments Use PTU 110A for 110 120V and PTU 220 for 220 240V operation 2 4 ATTACHING WATER TUBES The Pump and Tank Unit is supplied with two water tubes All connections are self sealing when disconnected to prevent water spills To connec...

Page 5: ...ion To protect the tip during shipment and when not in use a clear protective sheath is provided with the Thermal Probe The protective sleeve has a bayonet type locking feature and can be left in place on the probe shaft To expose the tip simply twist the sheath until the retaining screw is in the radial slot and slide the sheath back or forward along the probe shaft Twist it again to lock it in p...

Page 6: ...de exact setting All further temperature settings can be made using the digital control settings switches on the front panel These settings allow the temperature of the stage to be incremented or decremented in 5 1 degree or 1 10 steps from the base temperature 3 2 MEASURING ABSOLUTE TEMPERATURE The digital display of the controller may be used as a readout for an external sensor Type T copper con...

Page 7: ... of set point Safety shutdown with warning lamp in case of fault condition such as thermocouple sensor breakage lack of cooling water or electronics failure Self sealing water connectors Auxiliary AC switched output THERMAL PROBE SPECIFICATIONS SIZE 6 5 long x 5 8 dia Total length with cable 6 feet approx WEIGHT 0 5 lb TEMPERATURE CONTROLLED 0 5 diameter SURFACE AREA RANGE 0 45 C Note Sustained ex...

Page 8: ... this instrument call 973 779 5577 or e mail physitemp aol com In the event that any part of this system is to be returned for repair or recalibration please pack it with care and send it prepaid via Federal Express to Service Department PHYSITEMP INSTRUMENTS INC 154 Huron Avenue Clifton NJ 07013 U S A Please include with the instrument 1 A note describing any problems encountered 2 The name and t...

Reviews: